10 CALIBRATION
10.3 Calibration of scale
factor, ammeter 1
Scale factor, range LOW
1]
Press the menu-option
A
and change the
range to ”LOW”.
1]
Press
to confirm
1]
Press
A
to leave the menu.
2]
Connect a reference ammeter to the current
output.
Note:
If a current transformer is used, make
sure that it is applied at least 0.5 meter from
any bend of the current conductor. Other
conductors shall not be closer than 0.5 meter
from the current transformer.
3]
Press
HOLD
button to activate the Hold-func-
tion.
The lamp in the
HOLD
button lights up.
4]
Press the
MOM
button and adjust the current
until the calibration value (about 2/3 of maxi-
mal current, see chapter 11 Specifications,
section 11.2 -11.5) appears on the reference
ammeter. Release the
MOM
button when the
value is adjusted.
Note:
The lamp on the
HOLD
button should be
flashing.
5]
Press the
SYS
menu option.
6]
While simultaneously pressing the buttons
ESC
and
quickly turn the
Select/set knob
clockwise, until "CALIBRATION" appears on
the display.
7]
Press
to confirm.
8]
Choose “AMMETER 1”
9]
Press
.
10]
Turn the
Select/set knob
until the current
value on the display matches that on the
reference ammeter.
11]
Press
to confirm the calibration.
Scale factor, range HIGH
1]
Press the menu-option
A
and change the
range to ”HIGH”.
1]
Press
to confirm
1]
Press
A
to leave the menu.
2]
The rest of the procedure is the same as step
2 - 8 in the ”LOW”-range.
10.4 Scale factor for the
I/30-function
1]
See that the I/30-function is turned off.
2]
Activate the Hold-function by pressing
HOLD
(the LED in the button lights up).
3]
Press the
MOM
button and hold it down for a
second.
The LED in the
HOLD
button should start flash-
ing.
4]
Note the current value for Ammeter 1 that is
displayed on the display.
5]
Activate the I/30-function.
6]
Press the
MOM
button and hold it down for a
second.
The LED in the
HOLD
button should start flash-
ing.
7]
Press the
SYS
menu option.
8]
While simultaneously pressing the buttons
ESC
and
quickly turn the
Select/set knob
clockwise until "CALIBRATION" appears on
the display.
9]
Press
to confirm.
10]
Choose ”AMMETER 1”
11]
Press
.
”I/30” is displayed in the upper right corner
of the display.
12]
Turn the
Select/set knob
until the current
value on the display matches the value from
the previous measurement in step 4.
13]
Press
to confirm the calibration.
The I/30 function is intended only for use at coarse
adjustment of high currents and gives only an approxi-
mate indication.
Содержание INGVAR
Страница 2: ......
Страница 9: ...BH0654BE ZP BH05E 9 1 Safety...
Страница 17: ...BH0654BE ZP BH05E 17 2 Introduction...
Страница 19: ...BH0654BE ZP BH05E 19 4 Display...
Страница 33: ...BH0654BE ZP BH05E 33 5 Menu options...
Страница 47: ...BH0654BE ZP BH05E 47 5 Menu options...
Страница 67: ...BH0654BE ZP BH05E 67 Appendix 1...
Страница 70: ...70 ZP BH05E BH0654BE Index...
Страница 71: ......