STANDARD TEST PROCEDURE
AVTM246004 BITE 2/2P Rev 4 July 2011
43
Step Eight: What to Do When the Test Is Complete
You can perform the following operations after a test is
complete:
1.
Export test results to a PC. You can then use the PC to
view or print the test results. (Refer to Chapter 6.)
2.
Print test results on the BITE 2P
transmitter
printer.
3.
Delete the test information from the
receiver
.
4.
Start a new test. (Return to "Step One: Prepare for
Testing" page 27.)
5.
Power down and disconnect the
transmitter
. You will
need to do this if you are finished using the BITE 2/2P
for this testing session. (See "Powering Down and
Disconnecting the BITE 2/2P" on page 45).
Reviewing a Test
At any time while you are performing a test, you can review
the results of the active test that are already stored in the
receiver
.
1.
To review the current test, press
on the
receiver
to
scroll back through the active test screens.
2.
Press
to scroll forward through the active test screens.
You may print the active test results on the BITE 2P
transmitter
printer for review. Please note that there are no
statistics or bar graphs printed for the active test. See
"Printing an Active Test to Review the Data (BITE 2P)" on
page 44.
www.
GlobalTestSupply
.com
Find Quality Products Online at: