10-23.
Positive Relative Convergence
-
Purpose: operate Positive Relative Convergence test during
Subjective test
-
Chart: Near Vision Chart
-
Auxiliary Lens: None
-
Objective: measure the value of BLUR, BREAK, and RECOVERY
-
Operation sequence
1. Select horizontal Von Graefe chart
and then adjust
△X value to search blurred point.
2. After finding blurred point, touch [PRC] button on option controller
and then input the value on [Blur] and save it.
3. Change △X value again and search the break point. After finding,
input the value on [Break] and save it.
4.Readjust △X value in reverse to search recovery point. After finding,
input the value on [Recovery] and save it.