64
※ Attention
-
In case of distorted vision, this test is not available for the
normal visual acuity above 1.0 or an accurate visual acuity test.
-
Sometimes a patient does not respond to this test. In some
cases, a patient cannot compare the black letters in red and
green background respectively, and cannot react following his
preference.
-
If a corrected vision of a patient is 1.0, you must ask him
whether he is able to see well “29 in red and 92 in green.” In
case a corrected vision is low, ask him to read the bigger letter
on the upper side.
10-4.
Jackson Cross Cylinder Test
-
Purpose: Subjective test, revise maximum circumference refractive
power and axis for Monocular, using Jackson Cross Cylinder.
-
Chart: Dot-Group Chart
-
Auxiliary Lens: Cross Cylinder Lens (0.25 / 0.50)
-
Objective: The Dot-Group Chart must be seen as same, although the
Cross Cylinder is crossed.
-
Operation Sequence
1.
Tap (R) or (L) to choose the eye for test.
2.
Tap Dot-Group Chart button to move on to Cross Cylinder Test
Mode.
3.
Select Auxiliary Lens on option controller
4.
When you choose between [CC 0.25] and [CC 0.50], [1] and [2]
button appear at the bottom.