85
10-18.
Minute Stereo Acuity Test
-
Purpose: operate the minute stereo acuity test during subjective test.
-
Chart: Minute Stereo Vision Chart
-
Auxiliary Lens: polarized filter with 135 degree (for right eye), and
polarized filter with 45 degree (for left eye)
-
Objective: With fixation as a center, it must be seen as rotating
clockwise and as embossed.
-
Operation sequence
1.
Press minute stereo vision chart
and select minute
stereo vision test mode.
2.
Ask the patient, “How do you see the bars, with the fixation as a
center, from 12 to, 3, 6 and 9 o’clock direction? (clockwise)
-
All the bars look like flat (including central fixation): The
patient does not have the capability of stereo vision. Press
‘NG’ button.
-
The bars at 12 and 3 o’clock look like cubic: The patient can
perceive to 1 arc minute. Press [1°] button and setup it.
-
The bars at 3 and 6 o’clock look like cubic: The patient can
perceive to 2 arc minute. Press [2°] button and setup it.
-
The bars at 6 and 9 o’clock look like cubic: The patient can
perceive to 4 arc minute. Press [4°] button and setup it.
-
The bars at 12 and central Fixation look like cubic: The
patient can perceive to 10 arc minute. Press [10°] button
and setup it.
Содержание Smart DR13
Страница 1: ......
Страница 33: ...32 Fig 21 Controller View Screen...
Страница 37: ...36 Fig 25 List View and Detailed View Received data...
Страница 44: ...43 Fig 31 Set the Risley prism left and the Jackson cross cylinder Right...
Страница 46: ...45 Fig 33 Pop up selecting Auxiliary lens and CP Operation penal...
Страница 50: ...49 Fig 35 Adjustment of the SPH CYL AXIS Value...
Страница 61: ...60 Fig 50 Near Horizontal Von Graefe Test...
Страница 85: ...84...