USERS MANUAL |
25
Managing Leak Test Results Data:
Data can be managed, viewed, downloaded and printed
by using the following features:
Display Screen Data Review
- The display screen
features a procedure results review capability that
allows an operator or nurse the ability to scroll through
the past 2500 procedure results to confirm leak testing
protocol compliance, determine reprocessing activity
and review past results. Once 2500 tests have been
accumulated, each successive test replaces the oldest
of the prior 2500 tests. (see figure 20)
Accessing Past Leak Test Results
To access and review past procedure data follow
these steps:
• Press the
MENU
key on the keypad.
• Press the up/down arrows on the keypad until the
display arrow points to the “Stored Runs” selection.
• Press
CONTINUE
to select this option
• Press the “select run” option to access past test
data. Press
CONTINUE
to access stored runs.
• The last leak test run will appear at the top of the
screen (e.g. run 14). Use the up/down arrow keys
to select the desired leak test run. Press
CONTINUE
to view leak test result.
• When finished, press
CANCEL
to return to the list
of stored runs. Press
CANCEL
again to return to
the
MENU
.
• Press
CANCEL
to return to the original
START
menu.
Deleting Past Leak Test Results
To delete past procedure results follow these steps:
• Press the
MENU
key on the keypad.
• Press the up/down arrows on the keypad until the
display arrow points to the “Stored Runs” selection.
• Press
CONTINUE
to select this option
• Arrow down to select the “delete all” option and to
purge all past test data. To delete all stored runs,
press the
CONTINUE
button and the program will
begin the deletion process.
• When complete, the display screen will return to
the first
MENU
screen.
• Press
CANCEL
to return to the original
START
menu.
Other Menu Options
The
MENU
key on the keypad provides several types
of program tools:
Programming tools for the operator:
•
Set Date/Time:
A date/time feature allows the
Veriscan
system to time stamp every leak test.
See the System Setup section for directions on
setting the date/time.
•
Stored Runs:
The Veriscan system data storage
capability features the ability to store prior test
results. The Veriscan stores the last 2,500
leak tests.
•
Quick Reports:
The Veriscan system data
storage allows the user to print custom reports that
summarize daily, weekly and month leak testing
results. It also prints a list of all endoscopes that
failed a leak test within the previous 30 days. See the
Quick Report Printing area of the Data Access and
Management section for directions on printing Quick
Report data.
Programming tools for certified Veriscan system
technicians only:
These menu items should not be
accessed by any non-Veriscan system technician.
These technician-only settings include:
• System Status
• Self Test
• Sensor Cal
• Line Frequency
• Set Pressure UOM
• Pump/valve Control
NOTE:
Accessing these areas of the
MENU
program
will void the Medivators Reprocessing Systems
Veriscan Warranty. Contact Medivators Technical
Support at 1-800-444-4729 or your authorized
Medivators distributor if you have questions about this
area.