![Medivators VERISCAN LT Скачать руководство пользователя страница 17](http://html1.mh-extra.com/html/medivators/veriscan-lt/veriscan-lt_user-manual_1762519017.webp)
USERS MANUAL |
17
Veriscan’s “Screening Test” is an extremely accurate, dry endoscope leak test which is
ideal for evaluating “cold” or stable temperature endoscopes.
Use the “Screening Test” to:
1) Evaluate an endoscope that was not used on a patient within the last 30 minutes.
2) Evaluate a facility’s inventory of endoscopes.
Due to Veriscan’s extreme leak detection accuracy, it is recommended that all
endoscopes be tested within
a facility when a Veriscan unit is received for the first
time. A facility can also routinely test endoscopes using the “Screening Test” to
ensure ongoing early endoscope leak detection.
NOTE: The screening test is only to be used on stable temperature endoscopes
which have not been used within the last 30 minutes.
Performing the “Screening Test”
• Press the “Menu” button located on the keypad and the following
information will display:
Main Menu
Stored Runs
Quick Reports
Screening Test
Set Test Parameters
• Use the up/down arrow keys to align the curser with the “Screening Test” and then
press
“CONTINUE”
followed by
“START”
to begin the leak test.
• If optional “OPERATOR ID” screen activated, the following information
will display:
Screening Test Start
Enter Operator ID
Press “Continue”
Enter Operator ID number (maximum 4 digits) and then press
CONTINUE
.
• If optional “PATIENT ID” screen activated, the following information
will display:
Screening Test Start
Enter Patient ID
Press “Continue”
Enter Patient ID number (maximum 10 digits) and then press
CONTINUE
.
• The “ENTER SCOPE S/N” screen will now display as:
Screening Test Start
Enter Scope SN
Press “Continue”
Enter last three digits of endoscope’s serial number and then press
CONTINUE
(a
maximum of 10 digits can be entered).
• The leak test will now automatically begin when Veriscan is operating in the
factory default “
SCREENING TEST
” mode, and the following information will
be displayed:
Leak Test Start
Please standby...
Performing a Screening Test
Содержание VERISCAN LT
Страница 1: ...User Manual Automated Endoscope Leak Tester REPROCESSING SYSTEMS LT...
Страница 4: ......
Страница 35: ......