System Setup
Installation
• Position Veriscan system on a flat, level surface and within six (6) feet/
two (2) meters of the electrical AC wall outlet.
• Veriscan’s electrical requirements can be found under the “Introduction,
Product Specifications” section.
Attaching the Power Cord
•
WARNING:
Exercise caution when plugging in this electrical
device. Risk of electric shock.
• Co
nnect power cord to power cord outlet located on rear of unit. Gently
push cord in until fully inserted. (see figure 1)
• Plug the other end of the power cord into a standard AC wall outlet.
•
NOTE: This device must always be plugged into an electrical wall
outlet that includes an approved Ground Fault Interrupter (GFI)
prior to operation.
Powering the Veriscan System “ON”
• Press the power switch located on the back of the unit to turn system
power ‘on’. (see figure 2)
• When Veriscan is turned on, it will automatically perform an initialization
process which includes a series of diagnostic self-tests. When this
process is complete, the display will indicate that the unit is ready for
testing as displayed below:
mm/dd/yy hh:mm
Connect Scope
Press “Start”
Setting the Time and Date
The Veriscan’s internal clock will record the time of all leak test
procedures. To set the clock:
• Press the
MENU
button on the display keyboard
• Press the arrow down key until Set Date/Time selection is reached
• Press the
CONTINUE
key to select the Set Date/Time option
• To enter the month, press
CONTINUE
and then enter the correct month
in the “month = “ area: 1 = January, 2 = February, etc.. If an incorrect
entry is made, press
CANCEL
to
start the process over. Press
CONTINUE
when finished.
• To enter the date, press the down arrow key from the “month” row until
the “date” row is reached. Press
CONTINUE
to reach the date set
area. Enter the date. For single digit numbers, it is not necessary to
enter a “0” before the number (e.g. enter 3 for the 3
rd
). Press
CONTINUE
when finished.
• To enter the year, press the down arrow key from the “date” row until
the “year” row is reached. Press
CONTINUE
to reach the year set
area. Enter the year as a single digit where 6 = 2006, 70 = 2007, etc.
Press
CONTINUE
when finished.
Figure 1
Figure 2
Figure 3