USERS MANUAL |
23
• Use down arrow to toggle to “Yes” and press
CONTINUE
• Press the
CANCEL
button until the “Connect Scope” language is
displayed is on the screen.
Using MENU button for setting Automatic or Manual Test Mode
The Veriscan system is normally used in the Automatic Test Mode which uses factory
default parameter settings for endoscope leak testing. Veriscan always uses “Standard
Test 1” and “Screen Test 1” as the “Automatic” factory default parameter settings. If
special endoscope parameter settings are to be used for a leak test, then the Manual
Test Mode is to be used. The manual mode is only to be used for selecting alternative
endoscope parameter settings, other than the factory default “Standard Test” settings.
Alternative test parameter settings can affect Veriscan accuracy at detecting endoscope
leaks, and should only be performed by qualified and trained personnel. Alternative
parameter settings can be helpful when different settings are required for testing a
unique or specialized endoscope.
Activate the Automatic or Manual Test Mode by:
• Press the “MENU” button located on the keypad and the following information
will display:
Main Menu
Stored Runs
Quick Reports
Screening Test
Set Test Parameters
Use the up/down arrow keys to align the cursor with the “
Set Test Parameters
” and
then press “
CONTINUE
”.
• The “Set Test Parameters” screen will now display as:
Settings Mode: Auto or Manual (whichever active)
Use Oper ID: No (whichever active)
Use Pat ID: Yes (whichever active)
Use the up/down arrow keys to align the cursor with the “Setting Mode” and then
press “
CONTINUE
”.
• The “Select Settings Mode” screen will now display with “automatic” or “manual”
centered in the screen depending on which is active, plus:
Automatic (whichever active)
Automatic
Manual
Use the up/down arrow keys to align the cursor with “automatic” or “manual” mode
and then press
“CONTINUE
”. The selected mode is now saved.