| USERS MANUAL
Manual Test Mode
• If Veriscan is operating in the “manual” mode, the following information will display when performing
either the Standard or Screen leak test:
Standard Test
Select Test Setting
Standard Test 1
Standard Test 2
Standard Test 3
Screening Test
Select Test Setting
Screen Test 1
Screen Test 2
Screen Test 3
Up to three different endoscope parameter settings can be programmed and stored under the “Standard Test”
and also three under the “Screen Test”. Each parameter setting is stored as either 1, 2 or 3. Veriscan always
uses “Standard Test 1” and “Screen Test 1” as the factory default “Automatic” settings and it is not recommended
to change these settings. Any alternative endoscope parameter setting should be programmed and saved under
settings 2 or 3 for the Standard or Screen test. All settings are initially set to factory defaults, unless manually
programmed differently.
Contact Medivators Technical Support for further information.
NOTE: Veriscan’s leak detection accuracy is affected by parameter changes.
NOTE: Parameter changes should only be performed by Medivators trained and qualified personnel.
I
f the Veriscan is operating in the “Manual” test mode and needs to be changed back to the “Automatic” test
mode, then follow the steps listed under “
Activate the Automatic or Manual Test Mode by”
to select the
Automatic mode.
Contact Medivators Technical Support or your authorized Medivators distributor if further assistance is required.