Console Operation:
1. Set Up
The Set Up key function will allow you to enter patient data, set seat and pedal adjustments for
various knee ranges of motion and customize the settings of the MR100. When the Set Up key is
pressed the first option in the menu appears. Use the up/down arrows to scroll through the menu
and press the enter key to select an option.
Set Up menu:
1. Patient Data:
•
Age:
used in Vo2 and heart rate programs.
•
Gender:
used in Vo2 program.
•
Weight:
used in METs and Calorie calculations and Vo2 program.
•
Height:
used in the seat position and pedal crank set up program.
2. Seat Position:
•
User may input desired knee flexion angles (6 options) and the software will
calculate the seat’s fore/aft position and pedal position settings.
•
This feature is intended to aid in patient set up but may not be the final settings as
patient’s body symmetry may vary slightly.
•
This program uses the height from the Patient Data settings for limb length. The seat
back angle position is assumed to be in the center of the adjustment range. The six
knee angle options are:
1.
R Min
(Right leg minimum flexion)
L Min
(Left leg minimum flexion)
2.
R Max
(Right leg maximum flexion)
L Max
(Left leg maximum flexion)
3.
R Max
(Right leg maximum flexion)
L Min
(Left leg minimum flexion)
4.
R Min
(Right leg minimum flexion)
L Max
(Left leg maximum flexion)
5.
R Max
(Right leg maximum flexion)
R Min
(Right leg minimum flexion)
6.
L Max
(Left leg maximum flexion)
L Min
(Left leg minimum flexion)
3. Watts Per Row:
•
Adjusts the scale of the dot matrix when power (watts) is displayed. The default
setting is 10 watts per row. The default of 10 watts per row means the full display (all
10 rows lit) equal 100 watts.
•
The setting can be adjusted from 10 to 100 watts per row of lights on the graph.
4. Level Scale:
•
Set the amount of change in the level adjustment of workload (resistance at the
pedals) each time the arrow keys are pressed.
•
This feature allows you to have very fine increments of resistance for physically
challenged patients or set very high resistance levels for sports training. The default
setting is; Fine, 5 watts per level. The three options are:
1.
Fine
– 5 watts per level (at 60 rpm)
2.
Medium
– 10 watts per level (at 60 rpm)
3.
Coarse
– 15 watts per level (at 60 rpm)