Appendix A
45
/
46
or relocating the VS-10S/VS-10M/VS-10H video laryngoscope.
d Over the frequency range 150 kHz to 80 MHz, field strengths should be less than
[3] V/m
Recommended separation distances between portable and mobile RF communications
equipment and the VS-10S/VS-10M/VS-10H video laryngoscope
The VS-10S/VS-10M/VS-10H video laryngoscope is intended for use in an
electromagnetic environment in which radiated RF disturbances are controlled. The
customer or the user of the VS-10S/VS-10M/VS-10H video laryngoscope can help
prevent electromagnetic interference by maintaining a minimum distance between
portable and mobile RF communications equipment (transmitters) and the
VS-10S/VS-10M/VS-10H video laryngoscope as recommended below, according to
the maximum output power of the communications equipment.
Rated
maximum
output power
of transmitter
W
Separation distance according to frequency of transmitter m
150 kHz to 80
MHz
outside ISM
bands
150 kHz to 80
MHz
in ISM bands
80M~800MH
z
800M~2.7G
Hz
0.01
0.12
0.12
0.12
0.23
0.1
0.38
0.38
0.38
0.73
1
1.2
1.2
1.2
2.3
10
3.8
3.8
3.8
7.3
100
12
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended
P
d
2
.
1
P
d
2
.
1
P
d
2
.
1
P
d
3
.
2