Maintenance
31
/
46
9.1.3
Power Cable Check
Check the appearance of the power cable. If a surface damage or poor contact
between plug and socket is found, contact the distributor for replacement in
time.
Check whether the indicator of the video laryngoscope is illuminated when
connecting the power adapter to the power supply for battery charging. If the
video laryngoscope cannot be charged and cannot be started, contact the
distributor for maintenance in time.
9.1.4
Electrical Safety Test
To ensure safety, please conduct a dielectric strength test, leakage current test,
and ground impedance test according to the method stipulated in IEC60601-1.
9.2
Battery Maintenance
9.2.1
Battery Overview
The VS-10S/VS-10M/VS-10H video laryngoscope is equipped with a built-in
battery, which is charged through the power adapter. Before use, fully charge the
battery of the video laryngoscope to ensure continuous operation of the video
laryngoscope. It is not recommended that you use the video laryngoscope during
the charging. Use the built-in battery to power the video laryngoscope after
finishing charging the battery.
WARNING:
Improper use of the battery may shorten the service life of the battery.
Before using the device, check the battery to ensure that sufficient power is
available. Recharge the battery if required.