MEATEST
M631 Precision RTD Simulator
Operation manual
44
[:SOURce]:PLATinum:STANdard <CPD>
[:SOURce]:PLATinum:STANdard?
Description:
This command selects Platinum temperature standard.
Parameters:
<CPD>
{PT385A|PT385B|PT3916|PT3926|USER} (default PT385A)
·PT385A
Pt385 (68) standard
·PT385B
Pt385 (90) standard
·PT3916
Pt3916 standard
·PT3926
Pt3926 standard
·USER
User (see PLAT:COEF command)
Remarks:
Overlapped command
Value is set according to “Startup” preset on restart and to default on *RST command
Example:
PLAT:STAN PT385A
PLAT:STAN? Response: PT385A
[:SOURce]:PLATinum:ZRESistance
Syntax:
[:SOURce]:PLATinum:ZRESistance <DNPD>[<UNIT>]
[:SOURce]:PLATinum:ZRESistance?
Description:
This command sets resistance at 0 °C for Platinum function.
Parameters:
<DNPD>
Range 100.0 ... 1000.0 (default 100.0).
<UNIT>
{OHM}
·OHM
Remarks:
Overlapped command
Value is set according to “Startup” preset on restart and to default on *RST command
Example:
PLAT:ZRES 100.0
PLAT:ZRES? Response: 1.02 OHM
[:SOURce]:RESistance[:AMPLitude]
Syntax:
[:SOURce]:RESistance[:AMPLitude] <DNPD>[<UNIT>]
[:SOURce]:RESistance[:AMPLitude]?
Description:
This command sets amplitude in Resistance mode. Node SOUR:RES also selects
“RESISTANCE” function if not already selected. Optionally unit can be enclosed.
Parameters:
<DNPD>
Range 16.0 ... 400.0e3, default 100.0
<UNIT>
{OHM}
·OHM
Remarks:
Overlapped command
Value is set according to “Startup” preset on restart and to default on *RST command
Example:
RES 100.0
RES? Response: 1.02 OHM
[:SOURce]:TIMing:PCOunt?
Syntax:
[:SOURce]:TIMing:PCOunt?
Description: