MEATEST
M631 Precision RTD Simulator
Operation manual
26
Cable between decade and computer description (configuration 1:1)
Computer
D-Sub 1 D-Sub 2
M631
Receiver
2
2
Transmitter
Transmitter
3
3
Receiver
Ground
5
5
Ground
Table 3 RS232 cable connection
6.2.
GPIB Interface (option)
The simulator can be controlled via GPIB (General Purpose Interface Bus) interface.
Following equipment is required:
M631 simulator with LAN, USB, IEEE488 bus option
Personal Computer (or other controlling device) with GPIB interface
GPIB cable
The GPIB interface must be selected from simulator system menu to be in operation (
SETUP-
>Interface->Active bus
). There is only one GPIB setting accessible from the simulator system menu
under
SETUP->Interface
path:
GPIB Address
1 to 31
The instrument performs the following functions based on IEEE488 bus commands:
SH1, AH1, T5, L3, RL1, DC1
The instrument also recognizes the following general commands:
DCL Device Clear - resets the instrument to its basic state
SDC Selected Device Clear - resets the instrument to its basic state
GTL Go To Local - switches the remote control off
LLO Local Lock Out - switches the local control off, the instrument
cannot be controlled from the front panel
Commands are identical to the commands for RS-232 interface. Detailed
description is shown in chapter 8.2.
Figure 25 IEEE488 connector