MEATEST
M631 Precision RTD Simulator
Operation manual
16
Timing table
– a list of values in Ω and their durations in seconds. Browsing through the table is done
by ▲,▼ cursor keys. Editing is done using these softkeys:
Add
– create a new point.
Figure 16 Timing sequence point editing
Duration
– duration of selected resistance (from 2 ms to 10 000 s). Press the
SELECT
key
to proceed.
Resistance
– corresponding Ω value. Range is given by the actual resistance range of the
simulator.
Edit
– show the “Timing edit” panel for selected point to allow editing.
Delete
– delete selected point.
Cancel
– deactivate the Lookup Table and allow for curve saving.
Save
– closes the table and saves current settings.
Cancel
– closes the table and does not save current settings.
Editing an existing table
Existing table can be edited in the same way as it can be created. Editable entries (Preset name,
Timing table points) are selected using the
SELECT
key.
Figure 17 Timing sequence edit