MEATEST
M631 Precision RTD Simulator
Operation manual
28
6.4.
USB Interface (option)
The simulator can be controlled via USB (Universal Serial Bus) interface.
Following equipment is required:
M631 simulator with LAN, USB, IEEE488 bus option
Personal Computer (or other controlling device) with USB interface (USB type A
connector)
Standard USB A-B cable
The USB interface must be selected from simulator system menu to be in operation (
SETUP-
>Interface->Active bus
). There is no USB setting in the simulator.
Simulator is equipped with USB type B connector.
Pin
Label
Description
1
+5V
Power supply
2
DATA-
Data signal -
3
DATA+
Data
4
GND
Ground
Figure 28 USB connector
Communication from user control program is performed via standard RS232 interface. Following
settings should be set on your PC for proper operation:
Baudrate
9600 Bd
Data bits
8
Stop bits
1
Parity
None
Also proper COM port must be selected. After connecting simulator to your PC, virtual COM port
should appear in System Control panel of Microsoft Windows OS. This COM port is labeled “USB
Serial Port (COMxx)”.