M631 Precision RTD Simulator
MEATEST
19
Operation manual
4.7.
Main menu
Main
menu is displayed after pressing „Menu“ softkey. In main menu you can set all the parameters
available.
Figure 21 Main menu
Required menu item is highlighted using cursor keys ▲,▼ or display softkeys. Highlighted menu is
selected by pressing
SELECT
key or „Select“ softkey. Only editable items can be selected.
Information
This menu displays information about the device. Items can’t be changed by the user.
Manufacturer
Model
Serial number
Software version
Hardware version
Device
This menu permits setting operational parameters of device.
Temperature unit
Temperature functions can be expressed in different units. Available units are °C (Celsius), °F
(Fahrenheit) and K (Kelvin).
Switching
Item defines the way how resistance value is changed. Value R1 is changed to value R2 in the time
interval T. Resistance connected to the output terminals can have different values during the time
interval T.
FAST
Fastest possible switching method.
T is typically 400us.
Resistance during T is undefined. It can be higher or lower than R1 (R2).
SMOOTH
2 steps switching method with smallest possible resistance change.
T is typically 1ms.
VIA OPEN
2 steps switching method. R1 is switched to OPEN and than to R2.
T is typically 1ms.
VIA SHORT 2 steps switching method. R1 is switched to SHORT and than to R2.
T is typically 1ms.