Process Calibrator M505 (D)
MEATEST, s.r.o.
5
Operation manual v6
4
Graphics display
The LCD Display is divided into two parts. The upper part is for the multimeter functions, the lower part for the
calibrator. The bargraph in the middle is the analogical representation of the measured input signals. The left side
shows various symbols:
Symbols in the upper display part
AUTO
Automatic Voltage Range or firm measuring range
200V
Measure Ranges 2V, 20V, 200V, 100mA
INP
The display shows the input signal.
Symbols in the lower display part
RMP
Automatic Ramp of the calibrator signal
SUP
The battery is being charged
60%
Battery voltage in %. Do not run the instrument when the battery shows
0%
. Charge only with the
enclosed original charger.
5
Datalogger (Software Option)
The generated signal and the measured signal are stored in a table format. The date and the time are added. The
stored data can be downloaded via the USB to a PC. A Soft Manager supports the data transfer to a Windows PC.
Two Set Points
LEVEL LOW
and
LEVEL HIGH
and two Intervals INTERVAL 1 and INTERVAL 2 will be
selected. If the measured signal is between the both Set Points LEVEL LOW and LEVEL HIGH, the data will be
stored with
INTERVAL 1
. The
INTERVAL 2
stores the date when the input signal is outside the Set Points. The
both Intervals can be selected from 2 sec. to 24 hours.
With the key
LOGG
is the function Datalogger active. The key MENU and the two cursor keys permit the
parameter selection. With OK the setting will be stored.
MULTIMETER
measured input voltage in
automatic range.
CALIBRATOR
generates 12.000mA. The Battery
has 70% of capacity.
LOGGER OFF
Datalogger OFF
LOGGER ON
Data are stored
MEMORY ERASE
Erase the memory
INTERVAL 1:
< 2 Sek. … 24 Std.
INTERVAL 2:
< 2 Sek. … 24 Std.
LEVEL LOW:
free settable
LEVEL HIGH:
free settable