Process Calibrator M505 (D)
MEATEST, s.r.o.
13
Operation manual v6
The key
ESC
terminates the calibration and the display returns into the measuring mode.
11 Datalogger (model M505D only)
When the function Datalogger is activated, the calibrator output and the multimeter input signals will be stored in
the internal memory. This permits fast checking of transmitters and their calibration in the field.
Example: A T/C transmitter with mV input signal generates 0-10V at the output. Both signals are stored as a table
with date and time added from internal RTC.
The Data can be downloaded to the PC and displayed as tables and graphics. They can be used for Excel and
handled under Windows.
Set Points
LEVEL LOW
and
LEVEL HIGH
and Intervals
INTERVAL 1
and
INTERVAL 2
will be selected. If
the measured signal is within the set points LEVEL LOW and LEVEL HIGH, the data will be stored with
INTERVAL 1
. Outside the set points the Data are stored with
INTERVAL 2
. The Intervals are programmable
from 2 seconds to 24 hours.
DATA STORE
After LEVEL LOW and LEVEL HIGH and INTERVAL 1 and INTERVAL 2 are set, the data logging can be
initiated with
LOGGER ON
and
OK
.
In the menu step
LOGGER OFF
is the logging terminated.
The key MENU and the vertical cursors permit the selection of the parameters. They will be stored with the key
OK.
RECAL of stored DATA
The stored data can be uploaded to the PC via USB terminal. Enclosed
Softmanager M505W
supports the
communication.
Apart of the stored data, also the Transients cane be uploaded to the PC (see 5.2) and displayed as tables and
graphics and handled under Windows and Excel.
LOGGER OFF
Data logging disabled
LOGGER ON
Data logging enabled
MEMORY ERASE
Stored data will be erased
INTERVAL 1:
< 2 sec
… 24 h
INTERVAL 2:
< 2 sec
… 24 h
LEVEL LOW:
free selectable
LEVEL HIGH:
free selectable
Содержание M505
Страница 1: ...M505 D Process calibrator Operation manual ...
Страница 2: ......