Process Calibrator M505 (D)
MEATEST, s.r.o.
11
Operation manual v6
10.1 CALIBRATION of CALIBRATOR OUTPUTS
Select the required range (1) or (4).
Currents, Voltage, Resistors and Thermocouples will be calibrated in the Range (1).
In the Range (2) the Resistors and the RTD will be calibrated.
10.1.1 Calibration of Currents
Connect a mA-Meter to the calibrator outputs. Required accuracy 0.01% @ 25mADC
10.1.2 Calibration of Voltages
Connect V-Meter to the calibrator outputs. Required accuracy 0.01% @ 25VDC
Set the value with the keyboard until the connected mA-Meter
shows 4.000 mA.
SOURCE MODUS
.
Set the value with the keyboard until the connected mA-Meter
shows 20.000 mA.
SOURCE MODUS
.
Set the value with the keyboard until the connected mA-Meter
shows 4.000 mA.
SINK MODUS
.
Set the value with the keyboard until the connected mA-Meter
shows 20.000 mA.
SINK MODUS
.
Set the value with the keyboard until the connected V-Meter
shows 0.000 V.
Set the value with the keyboard until the connected V-Meter
shows 24.000 V.
Set the value with the keyboard until the connected V-Meter
shows 0.000 mV.
Set the value with the keyboard until the connected V-Meter
shows 500.000 mV.
Set the value with the keyboard until the connected V-Meter
shows 0.000 mV.
Set the value with the keyboard until the connected V-Meter
shows 25.000 mV.