Process Calibrator M505 (D) MEATEST, s.r.o.
Operation manual v7
10
9
Additional functions in the multimeter mode
The key
MENU
opens the Functions
TARA
and
FILTER
for the Multimeter:
ADDITIONAL FUNCTIONS
The numeric keyboard permits selection of following functions:
HOLD
Display Hold (with CLEAR return)
MAX
Maximum display reading of the Multimeter (with CLEAR return)
MIN
Minimum display reading of the Multimeter (with CLEAR return)
CLEAR
Cancel the above functions and switches into the measuring mode
10
Software – calibration
The voltage and the current ranges of the multimeter as well as the output signals of the calibrator can be calibrated
from the keyboard. The calibration menu is password protected. The calibration menu opens with the key
OK
pressed for 10 seconds.
Full calibration of all ranges or partial calibration of some selected ranges can be performed. A precision 5 digits
multimeter with accuracy of 0.005% is required.
Point 1: Calibrator Outputs
Points 2-3:
Multimeter Inputs
Point 4: Calibration of the Resistor Simulator
Point 5: Hardware Configuration
Point 6: Selection of Logo for customized instrument
Quit the Menu with
ESC
.
Please ask the manufacturer for the Password.
The TARE can be activated
The Filter can be set for the input signal
.