14) Enter the “@D2” to date stamp the printed
Data.
15) Enter the “@E” to terminate with another
end of line string.
The entire string must end with an ASCII 255. If
the final character is not followed by a 255
character “
º
” put one in. To force an ASCII
255 position the cursor 1 character past the last
wanted character and push the CLR key.
16) Finish out the String Entry by pushing
ENTER
twice.
Final String
å4åPEARSåå5@E@W3@D2@Eº
Push EXIT to return to normal scale operation
or use the UP/DOWN keys to continue through
the COMMPORT setup menus.
14
a
ALPHA
A
a
1
9
@
YZ
@
14
b
2
DEF
D
14
c
ALPHA
A
a
1
ALPHA
A
a
1
2
DEF
2
15
a
ALPHA
A
a
1
9
@
YZ
@
15
a
2
DEF
2
DEF
E
16
ENTER
ENTER
EXIT
SETUP
!
1)
Push
SETUP
followed immediately by the
key. The message reads “COM-
PORT1”.
2)
Push
ENTER
. The message reads “FOR-
MAT”. The “END LINE” entry screen is in
the FORMAT Menu.
➀
EXIT
SETUP
!
SETUP
➀
COMPORT1
➁
ENTER
FORMAT
PROGRAMMING THE END OF LINE STRING
The End of Line String is used to terminate print strings and is printer depen-
dent. The default string is Line Feed (LF=ASCII 10). In this example a Carriage
Return (CR=ASCII 13) will be added to the EOL String. Refer to the ASCII
table in the Appendix for proper codes.