MSI-3750 Digital Weight Indicator
◆
User Guide
Page 121
M E A S U R E M E N T S Y S T E M S I N T E R N A T I O N A L
To Completely Reset the 3750
This procedure will not work if the scale is already in the CAL SETUP Menus. Push the EXIT
key to get out of “CAL SETUP”.
1)
Turn off the 3750 Power. Expose the
Calibration Port by removing the seal screw
on the left side of the 3750 Meter housing.
2)
Insert a small non-metallic screwdriver or
the stem side of a wooden “Q-Tip” and
press the switch button in the hole. While
holding in the cal Switch, trun on the 3750.
The display reads “RESETALL” followed
by “R U SURE”.
3)
To complete RESETALL push
ENTER
.
To cancel this operation leaving the previous
calibration and setup registers intact, push
EXIT.
4)
The system will now be in the
“UNCAL’ED” state. Push the cal switch
and use the “INITIAL CALIBRATION”
procedure to calibrate the 3750.
➀
(see “Enable Calibration” for the location
of the Cal Port)
➁
POWER
➁
RESETALL
RU SURE?
③
ENTER
STD CAL
UNCAL’ED