Page 36
Control Commands
optoNCDT ILR 1181 / 1182
7.3.9
AWx.x......Display/Set ALARM Width
Standard setting: 100
- AW sets the length of the active range, beginning at AC.
- Set AW settings in agreement with the currently valid SF scale factor.
- AW is always equal or greater than “0” (zero).
- AW is always equal or greater than |AH| (the amount of AH), see Chap.
6.3
.
7.3.10 RBx.x.....Display/Set Distance of Lout=4mA
Standard setting: 0.1
RB (Range Beginning) corresponds to the starting point of the distance range that is provided at the analog
output.
- A distance value = RB will generate a current I
OUT
of 4 mA.
- Set RB in agreement with the currently valid SF scale factor.
- RB can be greater or smaller!
- Beyond the range that was set via RB and RE, the applied current will be that of the next limiting value.
In the event of a fault, the output value will correspond to the current that was set via parameter SE, see
Chap.
7.3.6
.
7.3.11 REx.x.....Display/Set Distance of Lout=20mA
Standard setting: 30
RE (Range End) corresponds to the end point of the distance range that is provided at the analog output.
- A distance value = RE will generate a current I
OUT
of 20 mA.
- Set RE in agreement of the scale factor SF.
- RE can be greater or smaller as RB!
- Beyond the range that was set via RB and RE, the applied current will be that of the next limiting value.
- In the event of a fault, the output value will correspond to the current that was set via parameter SE, see
Chap.
7.3.6
.
7.3.12 RMx y.y z......Remove Measurement
Standard setting: 0 0 0
RM is intended to facilitate settings for a range of expected distance values. Values which are found to be out-
side of this expected range will be corrected until matching the most recently valid measuring values.
The use of RM
parameter settings
should be restricted
to suitable applica-
tions only. Improper
use of this param-
eter may create
safety hazards!