PROGRAMMING EXAMPLE
Copyright © 2018 McCrometer, Inc. All printed material should not be changed or altered without permission of McCrometer.
Any published technical data and instructions are subject to change without notice. Contact your McCrometer representative
for current technical data and instructions.
www.mccrometer.com
3255 WEST STETSON AVENUE • HEMET, CALIFORNIA 92545 USA
Printed In The U.S.A.
TEL: 951-652-6811 • 800-220-2279 • FAX: 951-652-3078
Lit. # 30122-46 Rev. 1.2 / 5-31-18
25
7.0 PROGRAMMING EXAMPLE
The steps below demonstrate how to modify the full scale value from 3500 Gal/m to 3000 Gal/m on the “Quick Start
Menu”. You can enter the "Quick Start Menu" from any Visualization Page (see page 21 for information on this).
.....
.....
.....
.....
.....
.....
.....
.....
.....
.....
.....
.....
1. Long press on Enter/ESC.
This will wake up
the front panel
display.
4. Press Left/Right to find the next selection.
Press Left/Right
until you highlight
the
5
as shown.
2.
Short press on Enter/ESC.
This will take you
to the Quick Start
menu.
5.
Press Up/Down.
Press Up/Down to
change the value
to 0.
6. Short press on Enter/ESC.
This will confirm
and save the
setting.
7. Long press on Enter/ESC.
This will exit
programming
and return to the
visualization page.
You will start with a blank screen.
3.
Short press on Enter/ESC.
This will highlight
the GAL setting.
Note: The function used in the example is the top menu
item, so you will not need to scroll through the
Quick
Start
menu. However, to do so, you need to have the
full menu item highlighted and press
Left/Right
or
Up/
Down
.
Scroll up
Scroll down
Select