OPERATION
www.maxcessintl.com
MAGPOWR DLCA
MI 850A348 1 E
Tension calibration
Precision calibration
continued
Display now shows C1.5. Apply the calibration load to the
load cell using a temporary web or rope and a known weight
entered in C1.4. The temporaray web should follow the
normal web path. When using a rope, the rope should be
centered on the rollers. See Figure 5. Press [
] to start
calibration. The display will show a wait screen, and then
display PASS for a successful calibration or an error code to
indicate an issue with the calibration.
1 Temporary web
2 Sensing roll
3 Weight (W)
Figure 5. Apply calibration tension
From either the passing or error screen, press [
] to return
to displaying tension.
See the calibration errors in Table 3 on page 4-11.