7
Command reference
50
VCN? Command
This command allows receiving the setting value of output voltage with 0.0 to 100.0%
( increment of 0.01%)
(Syntax) VCN?
(Return)
①
VCN=100.0
Output voltage is set to a maximum rated voltage
②
VCN=12.34
Output voltage is set to 12.34% of maximum rated voltage
③
VCN=25.0
Output voltage is set to 25% of maximum rated voltage
ICN? Command
This command allows receiving the setting value of output current with 0.0 to 100.0%
( increment of 0.01%)
(Syntax) ICN?
(Return)
①
ICN=100.0
Output current is set to a maximum rated current
②
ICN=12.34
Output current is set to 12.34% of maximum rated current
③
ICN=25.0
Output current is set to 25% of maximum rated current
OVP? Command
This command allows receiving the setting value of over voltage protection setting value with 0.0 to
100.0%( increment of 0.1%)
Maximum over voltage protection value = maximum rated voltage x 110%
(Syntax) OVP?
(Return)
①
OVP=100.0
Set to a maximum over voltage protection value
②
OVP=12.3
Set to 12.3% of maximum over voltage protection value
③
OVP=25.0
Set to 25% of maximum over voltage protection value
OCP? Command
This command allows receiving the setting value of over current protection setting value with 0.0 to
100.0%( increment of 0.1%)
Maximum over current protection value = maximum rated current x 110%
(Syntax) OCP?
(Return)
①
OCP=100.0
Set to a maximum over current protection value
②
OCP=12.3
Set to 12.3% of maximum over current protection value
③
OCP=25.0
Set to 25% of maximum over current protection value