background image

Integrity Bath 

 

Model MB-80

 

Maintenance Manual 

  

07-06-15 
Page 5 of 16

 

2071 14

th

 Ave – Columbus, NE 68601 

800-798-5867  FAX: 402-563-9102 
www.mastercarebath.com

 

DOOR IN/OUT ADJUSTMENT 

Properly adjusted the door will stop just short of the 

inside 

tub rails.

 To adjust, locate the CAP HEAD SCREW that acts 

as a stop for the door. See item #

6

 coupled with item

 

#

4

 on 

drawing 

#804305.

 Loosen the LOCK NUT at the base of the 

screw. Adjust in and out as needed. Lock into place with 
LOCK NUT. See drawing to the right.

 

DOOR SLIDE END CAP 

Inspect the SLIDE END CAP item# 

3

 in drawing# 

803966

. It 

should not be worn through on the side that contacts the 

outer surface of the door. To replace: 

1.

 

Remove old end cap. 

2.

 

Clean any residual adhesive. 

3.

 

Smear  a  small  amount  of  clear  silicone  on  inside  of 
new cap. 

4.

 

Push on. 

DOOR GAS SPRING REPLACEMENT 

Release  door  by  pulling  out  on  door  catch  bracket  knob. 
The bracket on the door mount assembly (item #

5

 drawing 

804500

)  will  be  against  the  door  lock  mount  bracket. 

(item# 

10

  drawing 

804500)

.  Clamp  the  two  brackets 

together  using  a  cushion  between  the  clamp  terminal. 
Remove  the  hex  socket  pipe  plug  (item  #

1

  drawing

 

804500)

. Remove the old gas spring and replace with new 

spring, part #

802235

DRAWINGS 804500 

 

 

LOCK 
NUT 

CAP HEAD 
SCREW 

SLIDE END 
CAP 

Содержание MB-80

Страница 1: ...Integrity Bath Model MB 80 Maintenance Manual For All Maintenance Procedures Follow the Directions in this Manual...

Страница 2: ...o their application and the durability requirements of the healthcare industry It is recommended that the bathing system be inspected and preventative maintenance performed using the following schedul...

Страница 3: ...or and swing out of way Remove SOCKET HEAD CAP SCREWS from ROLLER BRACKET Save SS SPACERS between BRACKET and DOOR HINGE MOUNTING BASE b Add SS SPACER to each spacer group Reinsert SOCKET HEAD CAP SCR...

Страница 4: ...STIC WEDGE index is on top of the PLASTIC BUSHING b Position ALUMINUM MOUNTING BLOCK to insure even engagement of PLASTIC WEDGE Check the clearance of the bottom edge of the DOOR See drawing below You...

Страница 5: ...the SLIDE END CAP item 3 in drawing 803966 It should not be worn through on the side that contacts the outer surface of the door To replace 1 Remove old end cap 2 Clean any residual adhesive 3 Smear a...

Страница 6: ...bove directions Insert special tool in drain basket Item 12 slots Clamp on center of special tool with a pair of Vice Grips Remove by turning counter clockwise Clean old seal material away from drain...

Страница 7: ...itch Item 72 5 Disconnect the air tubing Item 79 from the switch nipple 6 Remove spade connectors from switch terminals noting which was connected to which terminal 7 Remove nut holding the switch in...

Страница 8: ...the replacing the cartridge 2 Using a utility knife gently pry off the cap 6 that is located on top of the thermostatic control valve handle 7 3 Using a Philips screwdriver remove the screw 8 and the...

Страница 9: ...0 798 5867 FAX 402 563 9102 www mastercarebath com GEN II MIXING VALVE DIAGRAM ITEM NO DESCRIPTION QTY 1 THERMOSTATIC VALVE 1 2 CARTRIDGE REPLACE THERMOSTATIC 1 3 O RING 3 4 HUB 1 5 SETTING RING 1 6 C...

Страница 10: ...d or cracked replace Check aeration body tightness against tub floor If loose tighten until 0 ring part 36000 under the shoulder seals With tub full of water and air on Check area around aeration gask...

Страница 11: ...Integrity Bath Model MB 80 Maintenance Manual 07 06 15 Page 11 of 16 2071 14 th Ave Columbus NE 68601 800 798 5867 FAX 402 563 9102 www mastercarebath com VARIABLE SPEED BLOWER ENTREE 500105...

Страница 12: ...with the pump operation will turn on when the pump operation is activated The middle Red LED is for Bath Oil The bottom Red LEO is for Shampoo Priming Operation Remove the empty one gallon container o...

Страница 13: ...ate Is the check valve inside the metering knob clogged To fix remove four screws to disassemble Inspect and clean ball spring and 0 ting Lubricate with liquid soap and reassemble Check the orifice to...

Страница 14: ...es but doesn t stay positioned when releasing lever Rotate clevis mount counter clockwise one turn and reinstall Rotate another turn if chair still doesn t lock in place Replacement 1 Support chair ba...

Страница 15: ...w it to shut off on its own 2 Turn the scale back on again by pushing the ON ZERO BUTTON After the scale turns back on push the ON ZERO BUTTON and continue to hold down while also pushing the LBS KG B...

Страница 16: ...ng Assembly Door Handle Assembly 301176 804220 Console Plumbing Assembly 804221 Console Plumbing Pre assembly 804225 Reservoir Assembly 804245 Drain Shoe Assembly Integrity Reservoir 804255 Plumbing A...