Mastercare MB-80 Скачать руководство пользователя страница 12

Integrity Bath 

 

Model MB-80

 

Maintenance Manual 

  

07-06-15 
Page 12 of 16

 

2071 14

th

 Ave – Columbus, NE 68601 

800-798-5867  FAX: 402-563-9102 
www.mastercarebath.com

 

DISPENSING SYSTEM 

If no solution is dispensed from the dispensing head, 
open the SERVICE DOOR to access the solutions 

storage. Check the fluid level in the one gallon 
BOTTLE. If it is below the intake hose, replace with 
MasterCare's Bath Oil or Shampoo per priming 
instructions below. 

Note:  

 

If you press Bath Oil or Shampoo button for 

more than five seconds, it will initiate a 
pump prime cycle. The normal pump cycle 
can finish and turn off before the five 
seconds occurs, but it will still enter Prime 
and turn back on. 

 

Holding a Button for a longer time will not 

cause any problems other than initiating 
Prime. 

 

 Pressing any button while a pump is 

running will shut off the pump. This is 
predominately a safety feature for .the 
Prime operation. 

 

Pressing multiple buttons at the same time 

will turn off any pumps and be otherwise 
ignored. 

 

A Red LED associated with the pump 

operation will turn on when the pump 
operation is activated. The middle Red LED 
is for Bath Oil. The bottom Red LEO is for 
Shampoo. 

Priming Operation: 

Remove the empty one gallon container of solutions and 
replace with MasterCare's Bath Oil or Shampoo. 
To prime, push appropriate button for the solution you are 
priming for seven seconds. 

SYSTEM DIAGNOSTICS: 

Pump does not dispense solution: 

 

Check fluid level in one gallon containers. If below 

bottom of intake hose, replace with MasterCare's Bath 
Oil or Shampoo.  

 

Check circuit breaker. lf tripped, re-set. If pump still 

does not operate, check all wiring connections. Call 1-
800-798-5867 for service tech support.  

 

 

ONE GALLON 

BOTTLES 

SERVICE DOOR 

OPENED 

BATH OIL 
BUTTON 

SHAMPOO 
BUTTON 

Caution:  Other  brand  bath  oils  and  shampoos 
may  be  used  with  the  MasterCare  System 
although  re-calibration  may  be  required. 

However, some products may not be compatible 
with  the  system  and  may  damage  components 
in  the  system,  thereby  voiding  the  warranty 
MasterCare  offers  an  extended  warranty  with 
your  use  of  MasterCare  bathing  and 
cleaning/disinfection  products.  Please  contact 
MasterCare  customer  service@  800-798-5867 

before  using  non-MasterCare  products  in  your 
system 

Содержание MB-80

Страница 1: ...Integrity Bath Model MB 80 Maintenance Manual For All Maintenance Procedures Follow the Directions in this Manual...

Страница 2: ...o their application and the durability requirements of the healthcare industry It is recommended that the bathing system be inspected and preventative maintenance performed using the following schedul...

Страница 3: ...or and swing out of way Remove SOCKET HEAD CAP SCREWS from ROLLER BRACKET Save SS SPACERS between BRACKET and DOOR HINGE MOUNTING BASE b Add SS SPACER to each spacer group Reinsert SOCKET HEAD CAP SCR...

Страница 4: ...STIC WEDGE index is on top of the PLASTIC BUSHING b Position ALUMINUM MOUNTING BLOCK to insure even engagement of PLASTIC WEDGE Check the clearance of the bottom edge of the DOOR See drawing below You...

Страница 5: ...the SLIDE END CAP item 3 in drawing 803966 It should not be worn through on the side that contacts the outer surface of the door To replace 1 Remove old end cap 2 Clean any residual adhesive 3 Smear a...

Страница 6: ...bove directions Insert special tool in drain basket Item 12 slots Clamp on center of special tool with a pair of Vice Grips Remove by turning counter clockwise Clean old seal material away from drain...

Страница 7: ...itch Item 72 5 Disconnect the air tubing Item 79 from the switch nipple 6 Remove spade connectors from switch terminals noting which was connected to which terminal 7 Remove nut holding the switch in...

Страница 8: ...the replacing the cartridge 2 Using a utility knife gently pry off the cap 6 that is located on top of the thermostatic control valve handle 7 3 Using a Philips screwdriver remove the screw 8 and the...

Страница 9: ...0 798 5867 FAX 402 563 9102 www mastercarebath com GEN II MIXING VALVE DIAGRAM ITEM NO DESCRIPTION QTY 1 THERMOSTATIC VALVE 1 2 CARTRIDGE REPLACE THERMOSTATIC 1 3 O RING 3 4 HUB 1 5 SETTING RING 1 6 C...

Страница 10: ...d or cracked replace Check aeration body tightness against tub floor If loose tighten until 0 ring part 36000 under the shoulder seals With tub full of water and air on Check area around aeration gask...

Страница 11: ...Integrity Bath Model MB 80 Maintenance Manual 07 06 15 Page 11 of 16 2071 14 th Ave Columbus NE 68601 800 798 5867 FAX 402 563 9102 www mastercarebath com VARIABLE SPEED BLOWER ENTREE 500105...

Страница 12: ...with the pump operation will turn on when the pump operation is activated The middle Red LED is for Bath Oil The bottom Red LEO is for Shampoo Priming Operation Remove the empty one gallon container o...

Страница 13: ...ate Is the check valve inside the metering knob clogged To fix remove four screws to disassemble Inspect and clean ball spring and 0 ting Lubricate with liquid soap and reassemble Check the orifice to...

Страница 14: ...es but doesn t stay positioned when releasing lever Rotate clevis mount counter clockwise one turn and reinstall Rotate another turn if chair still doesn t lock in place Replacement 1 Support chair ba...

Страница 15: ...w it to shut off on its own 2 Turn the scale back on again by pushing the ON ZERO BUTTON After the scale turns back on push the ON ZERO BUTTON and continue to hold down while also pushing the LBS KG B...

Страница 16: ...ng Assembly Door Handle Assembly 301176 804220 Console Plumbing Assembly 804221 Console Plumbing Pre assembly 804225 Reservoir Assembly 804245 Drain Shoe Assembly Integrity Reservoir 804255 Plumbing A...