Integrity Bath 

 

Model MB-80

 

Maintenance Manual 

  

07-06-15 
Page 8 of 16

 

2071 14

th

 Ave – Columbus, NE 68601 

800-798-5867  FAX: 402-563-9102 
www.mastercarebath.com

 

MIXING VALVE CARTRIDGE (ITEM #600578):

 

Note: 

Read

 

and follow all instructions thoroughly.

 

Refer to the GEN II Mixing Valve Diagram to reference details. Supplies 

needed: Silicone lubricant if replacing the cartridge (readily available at local hardware stores). 

Adjustment Instructions:  

1.

 

Using a utility knife, gently pry off the cap (6) that is located on top of the thermostatic control valve handle (7). 

2.

 

Using a Philips head screwdriver, remove the screw (8) and thermostatic control valve handle (7). 

3.

 

Orientation of the setting ring (5).

 

Note

:  When  the  two  red  lines  on  the  side  of  the  cartridge  (2)  line  up  the  water 

temperature  is  approximately  100

o

F.

  Rotate  the  setting  ring  (5)  counter  clockwise  a  notch  or  two  to  increase  the 

temperature. Doing so will increase the water temperature to approximately 110

o

F. 

4.

 

Replace the thermostatic control valve handle (7). Line up the stop on the handle (7) with the top right corner of the 
thermometer (refer to drawing). 

5.

 

Replace  the  screw  (8)  into  the  thermostatic  control  valve  handle  (7)  and  tighten.  Place  the  cap  (6)  into  the 
thermostatic control valve handle (7). 

Replacement Instructions: 

1.

 

Turn off the hot and cold water supply before the replacing the cartridge. 

2.

 

Using a utility knife, gently pry off the cap (6) that is located on top of the thermostatic control valve handle (7). 

3.

 

Using a Philips screwdriver, remove the screw (8) and thermostatic control valve handle (7). 

4.

 

Remove the escutcheon (9). Gently remove the hub (4) by carefully pulling off.  

5.

 

Remove the setting ring (5). 

6.

 

Using an adjustable wrench, loosen and remove the cartridge (2) from the thermostatic valve (1). 

7.

 

Apply a small amount of silicone lubricant to the three “O” rings (3) on the NEW cartridge (2). 

8.

 

Place the NEW cartridge (2) into the thermostatic valve (1) and hand tighten. Using an adjustable wrench, tighten the 
cartridge (2) accordingly (Note: DO NOT over tighten). 

9.

 

Install the setting ring (5) so it is horizontally positioned (3 and 9 o’clock position). 

Note

: When the two red lines on 

the  side  of  the  cartridge  (2)  line  up  the  water  temperature  is  approximately  100

o

F

.  To  increase  the  temperature, 

rotate  the  setting  ring  (5)  counter  clockwise  a  notch  or  two.  The  water  temperature  will  be  increased  to 
approximately 110

o

F. 

10.

 

 Replace the escutcheon (9) and hub (4) by gently pushing both into place. 

11.

 

Replace the thermostatic control valve handle (7). Line up the stop on the handle (7) with the top right corner of the 
thermometer (refer to drawing). 

12.

 

Turn on the water supply. Check the water temperature. If adjustment is needed, refer to step 9 accordingly. 

13.

 

Replace  the  screw  (8)  into  the  thermostatic  control  valve  handle  (7)  and  tighten.  Place  the  cap  (6)  into  the 

thermostatic control valve handle (7). 

 

 

Содержание MB-80

Страница 1: ...Integrity Bath Model MB 80 Maintenance Manual For All Maintenance Procedures Follow the Directions in this Manual...

Страница 2: ...o their application and the durability requirements of the healthcare industry It is recommended that the bathing system be inspected and preventative maintenance performed using the following schedul...

Страница 3: ...or and swing out of way Remove SOCKET HEAD CAP SCREWS from ROLLER BRACKET Save SS SPACERS between BRACKET and DOOR HINGE MOUNTING BASE b Add SS SPACER to each spacer group Reinsert SOCKET HEAD CAP SCR...

Страница 4: ...STIC WEDGE index is on top of the PLASTIC BUSHING b Position ALUMINUM MOUNTING BLOCK to insure even engagement of PLASTIC WEDGE Check the clearance of the bottom edge of the DOOR See drawing below You...

Страница 5: ...the SLIDE END CAP item 3 in drawing 803966 It should not be worn through on the side that contacts the outer surface of the door To replace 1 Remove old end cap 2 Clean any residual adhesive 3 Smear a...

Страница 6: ...bove directions Insert special tool in drain basket Item 12 slots Clamp on center of special tool with a pair of Vice Grips Remove by turning counter clockwise Clean old seal material away from drain...

Страница 7: ...itch Item 72 5 Disconnect the air tubing Item 79 from the switch nipple 6 Remove spade connectors from switch terminals noting which was connected to which terminal 7 Remove nut holding the switch in...

Страница 8: ...the replacing the cartridge 2 Using a utility knife gently pry off the cap 6 that is located on top of the thermostatic control valve handle 7 3 Using a Philips screwdriver remove the screw 8 and the...

Страница 9: ...0 798 5867 FAX 402 563 9102 www mastercarebath com GEN II MIXING VALVE DIAGRAM ITEM NO DESCRIPTION QTY 1 THERMOSTATIC VALVE 1 2 CARTRIDGE REPLACE THERMOSTATIC 1 3 O RING 3 4 HUB 1 5 SETTING RING 1 6 C...

Страница 10: ...d or cracked replace Check aeration body tightness against tub floor If loose tighten until 0 ring part 36000 under the shoulder seals With tub full of water and air on Check area around aeration gask...

Страница 11: ...Integrity Bath Model MB 80 Maintenance Manual 07 06 15 Page 11 of 16 2071 14 th Ave Columbus NE 68601 800 798 5867 FAX 402 563 9102 www mastercarebath com VARIABLE SPEED BLOWER ENTREE 500105...

Страница 12: ...with the pump operation will turn on when the pump operation is activated The middle Red LED is for Bath Oil The bottom Red LEO is for Shampoo Priming Operation Remove the empty one gallon container o...

Страница 13: ...ate Is the check valve inside the metering knob clogged To fix remove four screws to disassemble Inspect and clean ball spring and 0 ting Lubricate with liquid soap and reassemble Check the orifice to...

Страница 14: ...es but doesn t stay positioned when releasing lever Rotate clevis mount counter clockwise one turn and reinstall Rotate another turn if chair still doesn t lock in place Replacement 1 Support chair ba...

Страница 15: ...w it to shut off on its own 2 Turn the scale back on again by pushing the ON ZERO BUTTON After the scale turns back on push the ON ZERO BUTTON and continue to hold down while also pushing the LBS KG B...

Страница 16: ...ng Assembly Door Handle Assembly 301176 804220 Console Plumbing Assembly 804221 Console Plumbing Pre assembly 804225 Reservoir Assembly 804245 Drain Shoe Assembly Integrity Reservoir 804255 Plumbing A...

Отзывы: