Integrity Bath
–
Model MB-80
Maintenance Manual
07-06-15
Page 3 of 16
2071 14
th
Ave – Columbus, NE 68601
800-798-5867 FAX: 402-563-9102
www.mastercarebath.com
QUARTERLY INSPECTION: (
In addition to the monthly checks
the following should also be perfo1med quarterly
)
Tub System:
•
Dial thermometer
•
Aerators
•
Casual water drain
•
All fittings and connections to make sure they are secure
and no leaks have developed.
Instructions for inspection, adjustments and replacements for
the above items are under the section "Problem Solving And
Diagnosis."
lf deficiencies are found during any of the above inspections,
refer to the problem solving and diagnosis section of this
manual for instructions on how to correct.
SCHEDULE MAINTENANCE INSTRUCTIONS
DOOR SYSTEM
Door Roller and Cam Arm Adjustments & Maintenance
Maintenance Inspection:
a)
Inspect wear on ROLLER. If worn replace ROLLER
then shim if need. If not worn follow the
directions below.
b)
Reference drawing
803966 item #2
and
item #8
(2 cap screws) tighten quarterly.
To Increase Tension On The Door Seal:
a)
Open door and swing out of way. Remove
SOCKET HEAD CAP SCREWS from ROLLER
BRACKET. Save SS SPACERS between BRACKET
and DOOR HINGE MOUNTING BASE.
b)
Add SS SPACER to each spacer group. Reinsert
SOCKET HEAD CAP SCREWS through bracket and
spacers. Tighten securely with 3/16 Allen
Wrench.
Note: Normal (adequate) compression is achieved by
shimming cam roll bracket until door contacts seal and
continues to seal approximately 1/4". If door roller is over
shimmed accelerated wear on components will result.
Test the ground fault circuit interrupter protecting
this appliance periodically in accordance with the
manufacturer's instructions.
All deficiencies discovered during monthly and
quarterly inspections should be corrected and/or
repaired as soon as possible. This will help prevent
potential damage to your system that could occur if
corrective action is not taken and in the process,
extend the life of the system.
Failures of materials can occur due to normal wear
associated with use over time. Therefore, it is in the
best interests of the patient, staff and provider that
recommended safety inspections are scheduled on
a periodic basis and preventative maintenance
performed as needed.