background image

Integrity Bath 

 

Model MB-80

 

Maintenance Manual 

  

07-06-15 
Page 3 of 16

 

2071 14

th

 Ave – Columbus, NE 68601 

800-798-5867  FAX: 402-563-9102 
www.mastercarebath.com

 

QUARTERLY  INSPECTION:  (

In  addition  to  the  monthly  checks 

the following should also be perfo1med quarterly

Tub System: 

 

Dial thermometer 

 

Aerators 

 

Casual water drain 

 

All fittings and connections to make sure they are secure 

and no leaks have developed. 

Instructions for inspection, adjustments and replacements for 

the above items are under the section "Problem Solving And 
Diagnosis."  
lf deficiencies are found during any of the above inspections, 
refer  to  the  problem  solving  and  diagnosis  section  of  this 

manual for instructions on how to correct. 

SCHEDULE MAINTENANCE INSTRUCTIONS 
DOOR SYSTEM 

Door Roller and Cam Arm Adjustments & Maintenance 

Maintenance Inspection: 

a)

 

Inspect wear on ROLLER. If worn replace ROLLER 
then  shim  if  need.  If  not  worn  follow  the 
directions below. 

b)

 

Reference  drawing 

803966  item  #2

  and

  item  #8 

(2 cap screws) tighten quarterly. 

To Increase Tension On The Door Seal: 

a)

 

Open  door  and  swing  out  of  way.  Remove 
SOCKET  HEAD  CAP  SCREWS  from  ROLLER 
BRACKET.  Save  SS  SPACERS  between  BRACKET 

and DOOR HINGE MOUNTING BASE. 

b)

 

Add  SS  SPACER  to  each  spacer  group.  Reinsert 
SOCKET HEAD CAP SCREWS through bracket and 
spacers.  Tighten  securely  with  3/16  Allen 

Wrench. 

Note:  Normal  (adequate)  compression  is  achieved  by 
shimming  cam  roll  bracket  until  door  contacts  seal  and 
continues  to  seal  approximately  1/4".  If  door  roller  is  over 

shimmed accelerated wear on components will result. 

 

 

Test  the  ground  fault  circuit  interrupter  protecting 
this  appliance  periodically  in  accordance  with  the 
manufacturer's instructions.

 

All  deficiencies  discovered  during  monthly  and 
quarterly  inspections  should  be  corrected  and/or 
repaired as soon as possible. This will help prevent 

potential damage to your system that could occur if 
corrective  action  is  not  taken  and  in  the  process, 
extend the life of the system.

 

Failures of materials can occur due to normal wear 
associated with use over time. Therefore, it is in the 
best interests of the patient, staff and provider that 

recommended  safety  inspections  are  scheduled  on 
a  periodic  basis  and  preventative  maintenance 
performed as needed. 

SS SPACER 

ROLLER 

BRACKET 

ROLLER 

CAM 

DOOR ARM  

SOCKET HEAD 

CAP SCREWS 

Содержание MB-80

Страница 1: ...Integrity Bath Model MB 80 Maintenance Manual For All Maintenance Procedures Follow the Directions in this Manual...

Страница 2: ...o their application and the durability requirements of the healthcare industry It is recommended that the bathing system be inspected and preventative maintenance performed using the following schedul...

Страница 3: ...or and swing out of way Remove SOCKET HEAD CAP SCREWS from ROLLER BRACKET Save SS SPACERS between BRACKET and DOOR HINGE MOUNTING BASE b Add SS SPACER to each spacer group Reinsert SOCKET HEAD CAP SCR...

Страница 4: ...STIC WEDGE index is on top of the PLASTIC BUSHING b Position ALUMINUM MOUNTING BLOCK to insure even engagement of PLASTIC WEDGE Check the clearance of the bottom edge of the DOOR See drawing below You...

Страница 5: ...the SLIDE END CAP item 3 in drawing 803966 It should not be worn through on the side that contacts the outer surface of the door To replace 1 Remove old end cap 2 Clean any residual adhesive 3 Smear a...

Страница 6: ...bove directions Insert special tool in drain basket Item 12 slots Clamp on center of special tool with a pair of Vice Grips Remove by turning counter clockwise Clean old seal material away from drain...

Страница 7: ...itch Item 72 5 Disconnect the air tubing Item 79 from the switch nipple 6 Remove spade connectors from switch terminals noting which was connected to which terminal 7 Remove nut holding the switch in...

Страница 8: ...the replacing the cartridge 2 Using a utility knife gently pry off the cap 6 that is located on top of the thermostatic control valve handle 7 3 Using a Philips screwdriver remove the screw 8 and the...

Страница 9: ...0 798 5867 FAX 402 563 9102 www mastercarebath com GEN II MIXING VALVE DIAGRAM ITEM NO DESCRIPTION QTY 1 THERMOSTATIC VALVE 1 2 CARTRIDGE REPLACE THERMOSTATIC 1 3 O RING 3 4 HUB 1 5 SETTING RING 1 6 C...

Страница 10: ...d or cracked replace Check aeration body tightness against tub floor If loose tighten until 0 ring part 36000 under the shoulder seals With tub full of water and air on Check area around aeration gask...

Страница 11: ...Integrity Bath Model MB 80 Maintenance Manual 07 06 15 Page 11 of 16 2071 14 th Ave Columbus NE 68601 800 798 5867 FAX 402 563 9102 www mastercarebath com VARIABLE SPEED BLOWER ENTREE 500105...

Страница 12: ...with the pump operation will turn on when the pump operation is activated The middle Red LED is for Bath Oil The bottom Red LEO is for Shampoo Priming Operation Remove the empty one gallon container o...

Страница 13: ...ate Is the check valve inside the metering knob clogged To fix remove four screws to disassemble Inspect and clean ball spring and 0 ting Lubricate with liquid soap and reassemble Check the orifice to...

Страница 14: ...es but doesn t stay positioned when releasing lever Rotate clevis mount counter clockwise one turn and reinstall Rotate another turn if chair still doesn t lock in place Replacement 1 Support chair ba...

Страница 15: ...w it to shut off on its own 2 Turn the scale back on again by pushing the ON ZERO BUTTON After the scale turns back on push the ON ZERO BUTTON and continue to hold down while also pushing the LBS KG B...

Страница 16: ...ng Assembly Door Handle Assembly 301176 804220 Console Plumbing Assembly 804221 Console Plumbing Pre assembly 804225 Reservoir Assembly 804245 Drain Shoe Assembly Integrity Reservoir 804255 Plumbing A...

Отзывы: