background image

Integrity Bath 

 

Model MB-80

 

Maintenance Manual 

  

07-06-15 
Page 14 of 16

 

2071 14

th

 Ave – Columbus, NE 68601 

800-798-5867  FAX: 402-563-9102 
www.mastercarebath.com

 

CHAIR TRANSFER SYSTEM 

CHAIR TILT BACK GAS SPRING ADJUSTMENT 

(SAME DRAWING AS CURRENT MANUAL?? 

Adjustment:

 Reference drawing to right. 

1.

 

Support  chair  back  with  bungee  cord  to  seat  bottom  or 
have someone hold in upright position. 

2.

 

Remove  3/8-16  lock  nut  pivot  bolt  on  top  gas  spring  to 
allow  rotation  of  clevis  mount.  3)  With  clevis  slid  off 
mounting  boss  and  chair  back  with  jam  nut  loosened 
note the following: 

 

If  chair  back  will  not  recline:  Rotate  clevis  mount 

clockwise one turn and reinstall to test. Rotate one 
more if chair still doesn't recline.

 

 

If  chair  back  reclines  but  doesn't  stay  positioned 
when  releasing  lever:  Rotate  clevis  mount  counter 

clockwise one turn and reinstall. Rotate another turn 
if chair still doesn't lock in place

Replacement:

  

1.

 

Support  chair  back  same  as  above.  Remove  3/7-16  lock 
nut and top pivot bolt on top of gas spring. Grasp the gas 
spring on SS body turn counterclockwise to unscrew from 
gas spring extension. 

2.

 

Loosen jam nut on top of gas spring and note position of 
release lever to gas spring plunger. Unthread gas spring 
from  white  clevis  lever  mount  assembly  by  rotating  gas 
spring counter clockwise. 3) Install gas spring in reverse 

order. as follows: 

3.

 

Rotate gas spring clockwise into clevis mount until spring 
plunger  contacts  release  lever  with  lever  approximately 
horizontal.  Screw  gas  spring  lower  threads  into  spring 

extension. 

4.

 

Reinstall pivot bolt and lock nut. Test function. Refer to 
gas spring adjustment procedures above if adjustment is 

needed. 

 

 

Содержание MB-80

Страница 1: ...Integrity Bath Model MB 80 Maintenance Manual For All Maintenance Procedures Follow the Directions in this Manual...

Страница 2: ...o their application and the durability requirements of the healthcare industry It is recommended that the bathing system be inspected and preventative maintenance performed using the following schedul...

Страница 3: ...or and swing out of way Remove SOCKET HEAD CAP SCREWS from ROLLER BRACKET Save SS SPACERS between BRACKET and DOOR HINGE MOUNTING BASE b Add SS SPACER to each spacer group Reinsert SOCKET HEAD CAP SCR...

Страница 4: ...STIC WEDGE index is on top of the PLASTIC BUSHING b Position ALUMINUM MOUNTING BLOCK to insure even engagement of PLASTIC WEDGE Check the clearance of the bottom edge of the DOOR See drawing below You...

Страница 5: ...the SLIDE END CAP item 3 in drawing 803966 It should not be worn through on the side that contacts the outer surface of the door To replace 1 Remove old end cap 2 Clean any residual adhesive 3 Smear a...

Страница 6: ...bove directions Insert special tool in drain basket Item 12 slots Clamp on center of special tool with a pair of Vice Grips Remove by turning counter clockwise Clean old seal material away from drain...

Страница 7: ...itch Item 72 5 Disconnect the air tubing Item 79 from the switch nipple 6 Remove spade connectors from switch terminals noting which was connected to which terminal 7 Remove nut holding the switch in...

Страница 8: ...the replacing the cartridge 2 Using a utility knife gently pry off the cap 6 that is located on top of the thermostatic control valve handle 7 3 Using a Philips screwdriver remove the screw 8 and the...

Страница 9: ...0 798 5867 FAX 402 563 9102 www mastercarebath com GEN II MIXING VALVE DIAGRAM ITEM NO DESCRIPTION QTY 1 THERMOSTATIC VALVE 1 2 CARTRIDGE REPLACE THERMOSTATIC 1 3 O RING 3 4 HUB 1 5 SETTING RING 1 6 C...

Страница 10: ...d or cracked replace Check aeration body tightness against tub floor If loose tighten until 0 ring part 36000 under the shoulder seals With tub full of water and air on Check area around aeration gask...

Страница 11: ...Integrity Bath Model MB 80 Maintenance Manual 07 06 15 Page 11 of 16 2071 14 th Ave Columbus NE 68601 800 798 5867 FAX 402 563 9102 www mastercarebath com VARIABLE SPEED BLOWER ENTREE 500105...

Страница 12: ...with the pump operation will turn on when the pump operation is activated The middle Red LED is for Bath Oil The bottom Red LEO is for Shampoo Priming Operation Remove the empty one gallon container o...

Страница 13: ...ate Is the check valve inside the metering knob clogged To fix remove four screws to disassemble Inspect and clean ball spring and 0 ting Lubricate with liquid soap and reassemble Check the orifice to...

Страница 14: ...es but doesn t stay positioned when releasing lever Rotate clevis mount counter clockwise one turn and reinstall Rotate another turn if chair still doesn t lock in place Replacement 1 Support chair ba...

Страница 15: ...w it to shut off on its own 2 Turn the scale back on again by pushing the ON ZERO BUTTON After the scale turns back on push the ON ZERO BUTTON and continue to hold down while also pushing the LBS KG B...

Страница 16: ...ng Assembly Door Handle Assembly 301176 804220 Console Plumbing Assembly 804221 Console Plumbing Pre assembly 804225 Reservoir Assembly 804245 Drain Shoe Assembly Integrity Reservoir 804255 Plumbing A...

Отзывы: