background image

happen about 10 minutes, the meter will be turned 
off automatically. To turn it on, pushing the 

FUNC.

 

Button only. 

 

  REPLACING THE BATTERY   

 

                                       

WARNING

 

To  avoid  electrical  shock  or  personal 

injury,  remove  the  test  leads  and  any 

input  signals  before  replacing  the 

battery.  Replace  only  with  same  type 

of battery.

 

 
 
When the electrical tester displays the “

” mark 

or the backlight be not very lit, the battery must be 
replaced  to  maintain  proper  operation.  Use  the 
following procedure to replacing the battery: 

 

1. 

2. 

 

The  Rotary  Switch  is  used  to  select  OFF. 

Disconnect  test  leads  from  any  live  source 

and  remove  the  test  leads  from  the  input 

terminals. 

 18 

 

3. 

 

4. 

 

5. 

Remove  screws  on  the  battery  cover  and 
open the cover.

 

Remove  the  exhausted  battery  and  replace 
with three new 1.5V size AAA batteries. 
Place battery cover and secure by a screw. 

     

                                         

CAUTION 

Using this appliance in an environment 

with  a  strong  radiated  radio-frequency 

electromagnetic  field  (approximately 

3V/m)  may  influence  its  measuring 

accuracy.   

 

 

  ACCESSORIES   

 

   

Operator’s instruction manual 

 

    Set of test leads 

 

    Gift box 

 

    Battery 1.5V

3

  Size AAA   

 

 

 19 

 

Содержание MS2026R

Страница 1: ...d hand held current clamps with double insulation overvoltage category 1000V CAT II 600V CAT III and pollution 2 This meter complies with the requirements of the following European Community Directives 89 336 EEC Electromagnetic Compatibility and 73 23 EEC Low Voltage as amended by 93 68 EEC CE Marking However electrical noise or intense electromag netic fields in the vicinity of the equipment may...

Страница 2: ...and test leads for 2 any sign of damage or abnormality before every use If any abnormal conditions exist i e broken test leads cracked cases display not reading etc do not attempt to take any measurements Do not expose the instrument to direct sunlight extreme temperature or moisture Never ground yourself when taking electrical measurements Keep your body isolated from ground by using dry clothing...

Страница 3: ...inuity Test with battery operated There is the AC measurement to be RMS or TRUE RMS about the meter It is chosen as user buy 4 Otherwise there is a radius lamp near the clamp It is lighted in measuring current PANEL DESCRIPTION 1 2 3 Transformer jaws Pick up the AC current flowing through the conductor Rotary switch Rotary switch is used to select functions HOLD button When this button is pushed t...

Страница 4: ... to manual range push this button once When the meter operates in manual ranging mode push this button to change range to the higher one and hold this button for more than 3 seconds to return to auto range mode FUNC button In range press this button to select or or function Different symbol of function will appear on the LCD button Press the button is used to turn the backlight on Press the button...

Страница 5: ... touch any electrical conductors 3 4 The polarity of the red lead connection will be indicated along with the voltage value Read the measure result from the display AC VOLTAGE MEASUREMENT 1 Insert the black and red test leads into the COM and V Hz input terminals respectively 2 Set rotary switch at desired V position Connect the test lead tips in parallel with the circuit to be measured Be careful...

Страница 6: ...range indication OL will be display 10 2 When checking in circuit resistance be sure the circuit under test has all power removed and that all capacitors have been discharged fully CONTINUITY TESTING 1 2 3 4 Insert the black and red test leads into the COM and V Hz input terminals respectively Set rotary switch at desired position Push FUNC button to select Connect the test lead tips in parallel w...

Страница 7: ...witch to Hz position Connect the test leads across the source or load under measurement The polarity of the red lead connection is positive 12 4 Read the measure result from the display NOTE 1 2 The signal amplitude must also be greater than the sensitivity level Determine that the amplitude level of the signal to be measured is not greater than the input voltage limit 250V DC AC rms ADP MEASUREME...

Страница 8: ...5mm 86mm 32mm Weight Approx 330g Operating 5 to 35 Storage temperature 10 to 50 14 DC VOLTAGE Range Resolution Accuracy 6V 1mV 0 8 of rdg 5 digits 60V 10mV 600V 0 1V 1000V 1V 1 0 of rdg 2 digits Input Impedance 10M Overload Protection 1000V DC or 700V AC RMS AC VOLTAGE Range Resolution Accuracy 6V 1mV 1 2 of rdg 5 digits 60V 10mV 600V 0 1V 700V 1V Input Impedance 10M Frequency range 40Hz to 400Hz ...

Страница 9: ...0uF 1uF Overload Protection 250V dc or rms ac for all ranges 16 FREQUENCY Range Resolution Accuracy 40Hz 1Hz 0 1 of rdg 1digit 600Hz 1Hz 6kHz 1Hz 60kHz 10Hz 100kHz 0 1kHz Measurement range 1V to 10V rms 40Hz to 100kHz ADP Range Resolution Accuracy 600mV 0 1mV 3 0 of rdg 5 digits AUDIBLE CONTINUITY AND DIODE Range Description If continuity exists about less than 40 built in buzzer will sound Show t...

Страница 10: ...Use the following procedure to replacing the battery 1 2 The Rotary Switch is used to select OFF Disconnect test leads from any live source and remove the test leads from the input terminals 18 3 4 5 Remove screws on the battery cover and open the cover Remove the exhausted battery and replace with three new 1 5V size AAA batteries Place battery cover and secure by a screw CAUTION Using this appli...

Страница 11: ...OPERATOR S INSTRUCTION MANUAL CLAMP METER HYS005013 A1 ...

Отзывы: