s y s t e m   d e s c r i p t i o n

2-2

Rad-8 Signal Extraction Pulse Oximeter Operator’s Manual

Rad-8 Signal Extraction Pulse Oximeter Operator’s Manual

2-3

2

2

s y s t e m   d e s c r i p t i o n

Rad-8 Vertical

Rad-8 Horizontal

CONTROL / INDICATOR

DESCRIPTION

1

Power On / Off

Used to turn the unit on and off.

2

Up button
Down button

During saturation monitoring, use these buttons to adjust 
the volume of the pulse beep tone.

 Within the menu/setup system, these buttons are used to 
select values within each menu option.

NOTE:

 Pressing and holding down these buttons allow for 

the rapid scrolling of SpO2 and BPM alarm limits.

3

Next Button     

Used within the menu/setup system to move through setup 
options.  Not active during normal patient monitoring

4

Mode / Enter
 Button

Used to enter the setup menus and to select/activate cer-
tain entries within the menu/setup system.

5

Alarm Limits 
Button

Used to enter the alarm menu to adjust Hi/Low 

SpO

2

  

and 

heart rate alarm limits.

6

Alarm Silence 
Button

Push once to temporarily silence the alarm for 120 sec-
onds (default). Push a second time to return the unit to 
standard alarm monitoring after alarm condition has been 
corrected.

NOTE: 

The silence time can be set for 120, 90, 60 and 30 

seconds. See Section 4 - 

Setup Menu Level 1..

7

Pulse Rate 
Display

The pulse rate in beats per minute (bpm).  When search-
ing for a saturation and pulse, it will flash dashed lines.

R a d - 8   f r o n t   p a n e l   c o n t r o l s

CONTROL / INDICATOR

DESCRIPTION

8

Signal IQ /  
Pulse Bar

 The Signal IQ provides an indication of the quality of 
the acquired signal as well as the timing of the pulse.  
A green vertical LED bar rises and falls with the pulse, 
where the height of the bar indicates the quality of the 
signal.

9

Saturation 
Display

 The functional arterial hemoglobin oxygen saturation is 
displayed in units of SpO

2

. When searching for a satu-

ration and pulse, it will flash dashed lines.

10

Sensitivity 
Mode Button/
Indicator

Used to set the unit into Maximum Sensitivity, Normal 
Sensitivity, or APOD Mode.

11

FastSat Button/
Indicator

Used to set the unit in the FastSat Mode

12

Perfusion Index

The Perfusion Index provides an indication of the per-
centage of pulsatile signal to non pulsatile signal.  The 
bar is highest when the quality of the perfused site is 
best.

13

Trauma Button

Used to set the unit into the most sensitive mode and 
fastest averaging time.

14

Sensor Off 
Indicator

Sensor Off

Sensor off indicator will illuminate when the sensor is 
off the patient and an audible alarm will sound.

15

No Sensor 
Indicator

No Sensor

No sensor indicator will illuminate when no sensor is 
connected and an audible alarm will sound.

16

Battery Low 
Indicator

The yellow battery low indicator will illuminate with an 
audible alarm when the battery is low and requires 
recharging.

17

AC Power 
Charging 
Indicator

The AC Power Charging indicator is illuminated when 
the Rad-8 is connected to AC line power and the bat-
tery is charging.

18

Speaker

 Provides audible indication of alarm conditions, pulse 
tone and feedback for key-presses.  Ensure the speaker 
is not covered or the unit is placed face-down on bed-
ding or other sound absorbing surface.

19

Patient Cable 
Connector

Connects to a Direct Connect Spot Check (DCSC) sen-
sor or Masimo Patient Cable with a sensor. The icon 
next to the sensor indicates defibrillation proof.

10

8

9

6

5

4

2

3

1

11

12

7

14

15

13

19

17

16

18

1

4

5

3

8

9

13

10

16

14

15

6

17

11

7

12

2

19

18

Содержание Rad-8

Страница 1: ...Signal Extraction Pulse Oximeters OPERATOR S MANUAL...

Страница 2: ...9 297 7001 www masimo com EU Authorized Representative for Masimo Corporation MDSS GmbH Schiffgraben 41 30175 Hannover Germany Tel 49 511 62 62 86 30 Fax 49 511 62 62 86 33 MEDICAL ELECTRICAL EQUIPMEN...

Страница 3: ...If sensor is exposed to the irradiation the reading might be inaccurate or the unit might read zero for the duration of the active irradiation period Always remove the sensor from the patient and com...

Страница 4: ...ory default settings 4 10 Setup Menu Level 3 Clear trend 4 10 t a b l e o f c o n t e n t s iv Rad 8 Signal Extraction Pulse Oximeter Operator s Manual Rad 8 Signal Extraction Pulse Oximeter Operator...

Страница 5: ...the Rad 8 pulse oximeter appears before this introduction Other important safety information is located throughout the manual where appropriate Read the entire safety information section before you o...

Страница 6: ...ort 8 hours Internal battery life with fully charged battery Serial output port Display capability on Philips Agilent monitor through Philips VueLink function RadNet TM and RadLink capability INDICATI...

Страница 7: ...able to adapt to the varying physi ologic signals and or noise and separate them by looking at the whole signal and breaking it down to its fundamental components The Masimo SET signal processing alg...

Страница 8: ...he bottom of the Rad 8 vertical Rad 8 family offers full Masimo SET technology in a small compact device Rad 8 family supports the full line of Masimo sensors and patient cables see Section 8 Sensors...

Страница 9: ...rtical LED bar rises and falls with the pulse where the height of the bar indicates the quality of the signal 9 Saturation Display The functional arterial hemoglobin oxygen saturation is displayed in...

Страница 10: ...er or packaging and are defined below SYMBOLS RS 232 Equipotential Ground Terminal Caution consult accompanying documents Nurse Call Interface WEEE compliant Defibrillation Proof see front panel Mark...

Страница 11: ...er source CAUTION DO NOT CONNECT THE RAD 8 PULSE OXIMETER TO AN AC OUTLET CONTROLLED BY A SWITCH Verify the AC power voltage and frequency before use Verify that the power source can provide adequate...

Страница 12: ...e front panel display is free of alarm and system failure messages see Section 5 Alarms and Messages 9 Verify the display shows the following Mode setting Standard Std or Sleep SLP or Home Hnn SpO2 Lo...

Страница 13: ...DISABLED IN THIS MODE If the unit is connected to a RadNet system there will be no communication with RadNet in this mode 11 Verify that the patient alarms are functional by setting the high and low...

Страница 14: ...ity of the readings over time is affected by the averaging mode being used The longer the averaging time the more stable the readings tend to become This is due to a dampened response as the signal is...

Страница 15: ...rovides a visual indicator of the plethysmogram signal quality and an alert when the displayed SpO2 values are not based on adequate signal quality The signal quality indicator displayed on the Rad 8...

Страница 16: ...ound and can be silenced for 120 seconds default by pressing the Alarm Silence Button Refer to Setup Menu Level 1 in this section to change setting If a low battery condition occurs while not monitori...

Страница 17: ...rend memory while the device is turned on and the trend data remains in memory until the memory fills up or cleared by the user NOTE It is recommended that you clear the trend prior to performing a ne...

Страница 18: ...ing choose the preferred settings by stepping through menu options Refer to Section 5 Output 2 Connect one end of the VueLink cable to the Serial Output connector on the back of the Rad 8 3 Connect t...

Страница 19: ...e front panel of the oximeter The following sub sections describe each menu item in more detail The oximeter has options that allow user configuration to suit specific needs SPECIAL MENU STANDARD HOME...

Страница 20: ...Sensor 100 reserved 200 reserved 400 Low Signal IQ 800 Masimo SET This flag means the algorithm is running is full SET mode It requires a SET sensor and needs to acquire some clean data for this flag...

Страница 21: ...THE LIMITS EACH TIME THE PULSE OXIMETER IS USED An audible alarm and a flashing alarm status indicator will occur when an alarm limit is exceeded for greater than five seconds See section 4 Alarm Fea...

Страница 22: ...taneously Return for service 5 a l a r m s a n d m e s s a g e s MESSAGES The Rad 8 will indicate other data or system errors Message conditions for the Rad 8 follow DISPLAY TYPE SOLUTION LEDS FLASH H...

Страница 23: ...s Pulse Rate Accuracy Pulse rate 25 240 bpm No Motion1 Adults Pediatrics Neonate 3 digits Motion2 Adults Pediatrics Neonate 5 digits Low Perfusion3 Adults Pediatrics Neonates 3 digits Resolution Satur...

Страница 24: ...lass B Equipment Classification IEC 60601 1 UL 60601 1 IEC 60601 1 1 Type of Protection Class 1 on AC power Internally powered on battery power Degree of Protection Patient Cable Type BF Applied Part...

Страница 25: ...f monitoring For more information refer to the following table or contact your Sales Representative Use only Masimo SET sensors and sensor cables Select an appropriate sensor apply it as directed and...

Страница 26: ...o L 3 kg 3 3 3 bpm 5 bpm 3 3 bpm 40 kg 2 3 3 bpm 5 bpm 2 3 bpm LNCS NeoPt L 1 kg 3 3 3 bpm 5 bpm 3 3 bpm SENSOR APPLICATION INSTRUCTIONS Unless indicated otherwise in the directions for use reposition...

Страница 27: ...OLUTION DONOTSTER ILIZE ANY MASIMO SENSOR BY IRRADIATION STEAM OR ETHYLENE OXIDE M a s i m o S E T p a t i e n t c a b l e s Reusable patient cables of various lengths are available Only use appropria...

Страница 28: ...ot allow liquids to enter the interior of the instrument Glutaraldehyde Solution Ammonium Chloride Wipe 10 Chlorine bleach in H2O 70 Isopropyl alcohol CAUTIONS DO NOT AUTOCLAVE PRESSURE STERILIZE OR G...

Страница 29: ...ibed tests discontinue its use and correct the problem before returning the unit back to the user Before performing the following tests verify unit is connected to AC power Also disconnect any patient...

Страница 30: ...part of the price paid for this product MASIMO grants to Purchaser a nonexclusive nontransferable license without right to sublicense to use the copy of the incor porated software firmware and documen...

Страница 31: ...vative works based on the written materials without the prior written consent of MASIMO 6 Transfer Restrictions The software firmware is licensed to the Purchaser and may not be transferred to anyone...

Страница 32: ...tion on both the sensors and monitors to ensure accurate pulse oximetry when needed most 2007 Masimo Corporation Masimo Signal IQ Discrete SaturationTransform DST Satshare SET FastSat Rad 8 Radlink Fa...

Отзывы: