![Masimo Rad-8 Скачать руководство пользователя страница 23](http://html1.mh-extra.com/html/masimo/rad-8/rad-8_operators-manual_1735234023.webp)
Rad-8 Signal Extraction Pulse Oximeter Operator’s Manual
7-1
7
s p e c i f i c a t i o n s
R a d - 8 s p e c i f i c a t i o n s
PERFORMANCE
Measurement Range
SpO
2
:
1-100%
Pulse Rate:
25-240 beats per minute (bpm)
Perfusion:
0.02% - 20%
Response time:
<1 second delay
ACCURACY
Saturation
70% to 100%
No Motion
1
Adults,
Pediatrics
±
2 digits
Neonate
±
3 digits
Motion
2
Adults
1
, Pediatrics
1
±
3 digits
Neonate
±
3 digits
Low Perfusion
3
Adults,
Pediatrics
±
2 digits
Neonate
±
3 digits
Pulse Rate Accuracy
Pulse rate:
25-240 bpm
No Motion
1
Adults, Pediatrics, Neonate
±
3 digits
Motion
2
Adults, Pediatrics, Neonate
±
5 digits
Low Perfusion
3
Adults, Pediatrics, Neonates
±
3 digits
Resolution
Saturation (%SpO
2
)
1%
Pulse Rate (bpm)
1 bpm
ELECTRICAL
AC Power requirements:
100-240 VAC, 50-60 Hz
Power consumption:
15 VA max
.
Battery
Type:
Sealed lead acid
Capacity:
8
hours
4
Charging time:
8 hours
ENVIRONMENTAL
Operating Temperature:
41
°
F to 104
°
F (5
°
C to 40
°
C)
Transportation/Storage Temperature:
-40
°
F to 158
°
F (-40
°
C to +70
°
C)
5
Operating/Storage Humidity:
5% to 95%, non-condensing
Operating Altitude:
500 mbar to 1060 mbar pressure,
-1000 ft to 18,000 ft (-304 m to 5,486 m)