Rad-8 Signal Extraction Pulse Oximeter Operator’s Manual
8-3
8
s e n s o r s & p a t i e n t c a b l e s
LNOP
®
SPECIALTY SENSORS
(LNOP sensors must be used in conjunction with PC cables)
SENSOR
Weight
Saturation Accuracy
Pulse Rate Accuracy
Low Perfusion Accuracy
Range
No Motion
Motion No Motion
Motion
Saturation
Pulse Rate
LNOP Blue
2.5 - 30 kg
60 - 80% ± 4%
N/A
± 3 bpm
N/A
± 3%
± 3 bpm
70 - 100% ± 3.3%
N/A
± 3 bpm
N/A
± 3%
± 3 bpm
80 - 100% ± 3%
N/A
± 3 bpm
N/A
± 3%
± 3 bpm
LNOP
Newborn Neo
< 3 kg
± 3%
± 3%
± 3 bpm
± 5 bpm
± 3%
± 3 bpm
LNOP
Newborn Inf
3 - 20 kg
± 2%
± 3%
± 3 bpm
± 5 bpm
± 2%
± 3 bpm
LNOPv
TM
ADHESIVE SENSORS
(LNOPv sensors must be used in conjunction with PC cables)
SENSOR
Weight
Range
Saturation Accuracy
Pulse Rate Accuracy
Low Perfusion Accuracy
No Motion
Motion
No Motion
Motion
Saturation
Pulse Rate
LNOPv In
3 - 20 kg
± 2%
± 3%
± 3 bpm
± 5 bpm
± 2%
± 3 bpm
LNOPv Ne
< 3 kg
± 3%
± 3%
± 3 bpm
± 5 bpm
± 3%
± 3 bpm
LNOPv Ad
> 30 kg
± 2%
± 3%
± 3 bpm
± 5 bpm
± 2%
± 3 bpm
LNCS
TM
REUSABLE SENSORS
(LNCS sensors must be used in conjunction with LNC cables)
SENSOR
Weight
Range
Saturation Accuracy
Pulse Rate Accuracy
Low Perfusion Accuracy
No Motion
Motion
No Motion
Motion
Saturation
Pulse Rate
LNCS DCI
> 30 kg
± 2%
± 3%
± 3 bpm
± 5 bpm
± 2%
± 3 bpm
LNCS DCIP
10 - 50 kg
± 2%
± 3%
± 3 bpm
± 5 bpm
± 2%
± 3 bpm
LNCS TC-I
> 30 kg
± 3.5%
N/A
± 3 bpm
N/A
± 3.5%
± 3 bpm
LNCS TF-I
> 30 kg
± 2%
N/A
± 3 bpm
N/A
± 2%
± 3 bpm
LNCS YI
> 1 kg
± 2%
± 3%
± 3 bpm
± 5 bpm
N/A
N/A
NOTE:
The LNCS TF-I and TC-I sensors were not validated under motion conditions.
LNCS
TM
ADHESIVE SENSORS
(LNCS sensors must be used in conjunction with LNC cables)
SENSOR
Weight
Range
Saturation Accuracy
Pulse Rate Accuracy
Low Perfusion Accuracy
No Motion
Motion
No Motion
Motion
Saturation
Pulse Rate
LNCS Adtx
> 30 kg
± 2%
± 3%
± 3 bpm
± 5 bpm
± 2%
± 3 bpm
LNCS Pdtx
10 - 50 kg
± 2%
± 3%
± 3 bpm
± 5 bpm
± 2%
± 3 bpm
LNCS Inf-L
3 - 20 kg
± 2%
± 3%
± 3 bpm
± 5 bpm
± 2%
± 3 bpm
LNCS Neo-L
< 3 kg
± 3%
± 3%
± 3 bpm
± 5 bpm
± 3%
± 3 bpm
> 40 kg
± 2%
± 3%
± 3 bpm
± 5 bpm
± 2%
± 3 bpm
LNCS NeoPt-L
< 1 kg
± 3%
± 3%
± 3 bpm
± 5 bpm
± 3%
± 3 bpm
SENSOR APPLICATION INSTRUCTIONS
Unless indicated otherwise in the directions for use, reposition reusable sensors at least
every 4 hours and for adhesive sensors inspect the site at least every 8 hours or sooner. If
indicated by circulatory condition or skin integrity, reapply to a different monitoring site.
LNOP
®
DIRECT CONNECT REUSEABLE SENSORS
SENSOR
Weight
Range
Saturation Accuracy
Pulse Rate Accuracy
Low Perfusion Accuracy
No Motion
Motion
No Motion
Motion
Saturation
Pulse Rate
LNOP DCSC
> 30 kg
± 2%
± 3%
± 3 bpm
± 5 bpm
± 2%
± 3 bpm
LNOP DC-12
> 30 kg
± 2%
± 3%
± 3 bpm
± 5 bpm
± 2%
± 3 bpm
LNOP
®
REUSABLE SENSORS
(
LNOP sensors must be used in conjunction with PC cables)
SENSOR
Weight
Range
Saturation Accuracy
Pulse Rate Accuracy
Low Perfusion Accuracy
No Motion
Motion
No Motion
Motion
Saturation
Pulse Rate
LNOP DCI
> 30 kg
± 2%
± 3%
± 3 bpm
± 5 bpm
± 2%
± 3 bpm
LNOP DCIP
10 - 50 kg
± 2%
± 3%
± 3 bpm
± 5 bpm
± 2%
± 3 bpm
LNOP YI
> 1 kg
± 2%
± 3%
± 3 bpm
± 5 bpm
N/A
N/A
LNOP TC-I
> 30 kg
± 3.5%
N/A
± 3 bpm
N/A
± 3.5%
± 3 bpm
LNOP DC-195
> 30 kg
± 2%
± 3%
± 3 bpm
± 5 bpm
± 2%
± 3 bpm
LNOP TF-I
> 30 kg
± 2%
N/A
± 3 bpm
N/A
± 2%
± 3 bpm
NOTE:
The LNOP TF-I and TC-I sensors were not validated under motion conditions.
LNOP
®
ADHESIVE SENSORS
(
LNOP sensors must be used in conjunction with PC cables)
SENSOR
Weight
Range
Saturation Accuracy
Pulse Rate Accuracy
Low Perfusion Accuracy
No Motion
Motion
No Motion
Motion
Saturation
Pulse Rate
LNOP Adt
> 30 kg
± 2%
± 3%
± 3 bpm
± 5 bpm
± 2%
± 3 bpm
LNOP Adtx
> 30 kg
± 2%
± 3%
± 3 bpm
± 5 bpm
± 2%
± 3 bpm
LNOP Pdt
10 - 50 kg
± 2%
± 3%
± 3 bpm
± 5 bpm
± 2%
± 3 bpm
LNOP Pdtx
10 - 50 kg
± 2%
± 3%
± 3 bpm
± 5 bpm
± 2%
± 3 bpm
LNOP Neo
< 10 kg
± 3%
± 3%
± 3 bpm
± 5 bpm
± 3%
± 3 bpm
LNOP NeoPt
< 1 kg
± 3%
± 3%
± 3 bpm
± 5 bpm
± 3%
± 3 bpm
LNOP Neo-L
< 3 kg
± 3%
± 3%
± 3 bpm
± 5 bpm
± 3%
± 3 bpm
> 40 kg
± 2%
± 3%
± 3 bpm
± 5 bpm
± 2%
± 3 bpm
LNOP NeoPt-L
< 1 kg
± 3%
± 3%
± 3 bpm
± 5 bpm
± 3%
± 3 bpm
LNOP Inf-L
3 - 20 kg
± 2%
± 3%
± 3 bpm
± 5 bpm
± 2%
± 3 bpm
s e n s o r s & p a t i e n t c a b l e s
8-2
Rad-8 Signal Extraction Pulse Oximeter Operator’s Manual
8