Masimo Rad-8 Скачать руководство пользователя страница 3

The Rad-8 Compact pulse oximeter is designed to minimize the possibility of hazards from 
errors in the software program by following sound engineering design processes, risk analysis 
and software validation.

 

Explosion hazard.  Do not use the pulse oximeter in the presence of flammable anes-
thetics or other flammable substance in combination with air, oxygen-enriched environ-
ments, or nitrous oxide.

 

High intensity extreme lights (such as pulsating strobe lights) directed on the sensor, 
may not allow the pulse oximeter to obtain vital sign readings.

 

The pulse oximeter is NOT intended for use as an apnea monitor.

 

A pulse oximeter should be considered an early warning device.  As a trend towards 
patient hypoxemia is indicated, blood samples should be analyzed by laboratory 
instruments to completely understand the patient’s condition.

 

The pulse oximeter is to be operated by qualified personnel only.  This manual, acces-
sory directions for use, all precautionary information, and specifications should be 
read and understood before use.

 

Electric shock hazard. Do not open the pulse oximeter cover except to replace the 
battery of the unit. Only a qualified operator may perform maintenance procedures 
specifically described in this manual. Refer servicing to Masimo for repair of this 
equipment.

 

Only a qualified operator may perform maintenance procedures specifically described 
in this manual.  Refer servicing to Masimo for repair of this equipment.

 

As with all medical equipment, carefully route patient cabling to reduce the possibility 
of patient entanglement or strangulation.

 

Do not place the pulse oximeter or accessories in any position that might cause it to 
fall on the patient.  Do not lift the pulse oximeter by the patient cable.  

 

Interfering Substances: Carboxyhemoglobin and Methemoglobin may erroneously 
increase readings.  The level of increase is approximately equal to the amount of car-
boxyhemoglobin present.  Dyes, or any substance containing dyes, that change usual 
blood pigmentation may cause erroneous readings.

 

Severe anemia may cause erroneous SpO

2

 readings.

 

Do not use the pulse oximeter or oximetry sensors during magnetic resonance 
imaging (MRI) scanning.  Induced current could potentially cause burns.  The pulse 
oximeter may affect the MRI image, and the MRI unit may affect the accuracy of the 
oximetry measurements.

 

If using pulse oximetry during full body irradiation, keep the sensor out of the irradiation 
field.  If sensor  is exposed to the irradiation, the reading might be inaccurate or the unit 
might read zero for the duration of  the active irradiation period.

 

Always remove the sensor from the patient and completely disconnect the patient 
from the pulse oximeter before bathing the patient.

 

Do not place the pulse oximeter where the controls can be changed by the patient.

 

Do not place the pulse oximeter face against a surface.  This will cause the alarm to 
be muffled.

ii

Rad-8 Signal Extraction Pulse Oximeter Operator’s Manual

Rad-8 Signal Extraction Pulse Oximeter Operator’s Manual

iii

 

Do not place the pulse oximeter on electrical equipment that may affect the pulse 
oximeter, preventing it from working properly.

 

Do not expose the pulse oximeter to excessive moisture such as direct exposure to 
rain.  Excessive moisture can cause the pulse oximeter to perform inaccurately or fail.

 

Do not place containers containing liquids on or near the pulse oximeter.  Liquids 
spilled on the pulse oximeter may cause it to perform inaccurately or fail.

 

Failure of Operation - If the pulse oximeter fails any part of the setup procedures remove the 
pulse oximeter from operation until qualified service personnel have corrected the situation.

 

Patient Safety - If a sensor or cable is damaged in any way, discontinue use 
immediately.

 

Disposal of product - Comply with local laws in the disposal of the unit and/or its 
accessories.

■ 

The pulse oximeter can be used during defibrillation, but the readings may be inac-
curate for up to 20 seconds.

 

This equipment has been tested and found to comply with the limits for medical 
devices to the EN 60601-1-2: 2002, Medical Device Directive 93/42/EEC.  These limits 
are designed to provide reasonable protection against harmful interference in a typical 
medical installation.  This equipment generates, uses and can radiate radio frequency 
energy and, if not installed and used in accordance with the instructions, may cause 
harmful interference to other devices in the vicinity.  However, there is no guarantee 
that interference will not occur in a particular installation.  If this equipment does cause 
harmful interference to other devices, which can be determined by turning the equip-
ment off and on, the user is encouraged to try to correct the interference by one or 
more of the following measures:

   Reorient or relocate the receiving device.

■  

Increase the separation between the equipment.

■  

Consult the manufacturer for help.

 

A functional tester cannot be utilized to assess the accuracy of the pulse oximeter or 
any sensors.

SAFETY INFORMATION, WARNINGS, CAUTIONS AND NOTES

SAFETY INFORMATION, WARNINGS, CAUTIONS AND NOTES

Содержание Rad-8

Страница 1: ...Signal Extraction Pulse Oximeters OPERATOR S MANUAL...

Страница 2: ...9 297 7001 www masimo com EU Authorized Representative for Masimo Corporation MDSS GmbH Schiffgraben 41 30175 Hannover Germany Tel 49 511 62 62 86 30 Fax 49 511 62 62 86 33 MEDICAL ELECTRICAL EQUIPMEN...

Страница 3: ...If sensor is exposed to the irradiation the reading might be inaccurate or the unit might read zero for the duration of the active irradiation period Always remove the sensor from the patient and com...

Страница 4: ...ory default settings 4 10 Setup Menu Level 3 Clear trend 4 10 t a b l e o f c o n t e n t s iv Rad 8 Signal Extraction Pulse Oximeter Operator s Manual Rad 8 Signal Extraction Pulse Oximeter Operator...

Страница 5: ...the Rad 8 pulse oximeter appears before this introduction Other important safety information is located throughout the manual where appropriate Read the entire safety information section before you o...

Страница 6: ...ort 8 hours Internal battery life with fully charged battery Serial output port Display capability on Philips Agilent monitor through Philips VueLink function RadNet TM and RadLink capability INDICATI...

Страница 7: ...able to adapt to the varying physi ologic signals and or noise and separate them by looking at the whole signal and breaking it down to its fundamental components The Masimo SET signal processing alg...

Страница 8: ...he bottom of the Rad 8 vertical Rad 8 family offers full Masimo SET technology in a small compact device Rad 8 family supports the full line of Masimo sensors and patient cables see Section 8 Sensors...

Страница 9: ...rtical LED bar rises and falls with the pulse where the height of the bar indicates the quality of the signal 9 Saturation Display The functional arterial hemoglobin oxygen saturation is displayed in...

Страница 10: ...er or packaging and are defined below SYMBOLS RS 232 Equipotential Ground Terminal Caution consult accompanying documents Nurse Call Interface WEEE compliant Defibrillation Proof see front panel Mark...

Страница 11: ...er source CAUTION DO NOT CONNECT THE RAD 8 PULSE OXIMETER TO AN AC OUTLET CONTROLLED BY A SWITCH Verify the AC power voltage and frequency before use Verify that the power source can provide adequate...

Страница 12: ...e front panel display is free of alarm and system failure messages see Section 5 Alarms and Messages 9 Verify the display shows the following Mode setting Standard Std or Sleep SLP or Home Hnn SpO2 Lo...

Страница 13: ...DISABLED IN THIS MODE If the unit is connected to a RadNet system there will be no communication with RadNet in this mode 11 Verify that the patient alarms are functional by setting the high and low...

Страница 14: ...ity of the readings over time is affected by the averaging mode being used The longer the averaging time the more stable the readings tend to become This is due to a dampened response as the signal is...

Страница 15: ...rovides a visual indicator of the plethysmogram signal quality and an alert when the displayed SpO2 values are not based on adequate signal quality The signal quality indicator displayed on the Rad 8...

Страница 16: ...ound and can be silenced for 120 seconds default by pressing the Alarm Silence Button Refer to Setup Menu Level 1 in this section to change setting If a low battery condition occurs while not monitori...

Страница 17: ...rend memory while the device is turned on and the trend data remains in memory until the memory fills up or cleared by the user NOTE It is recommended that you clear the trend prior to performing a ne...

Страница 18: ...ing choose the preferred settings by stepping through menu options Refer to Section 5 Output 2 Connect one end of the VueLink cable to the Serial Output connector on the back of the Rad 8 3 Connect t...

Страница 19: ...e front panel of the oximeter The following sub sections describe each menu item in more detail The oximeter has options that allow user configuration to suit specific needs SPECIAL MENU STANDARD HOME...

Страница 20: ...Sensor 100 reserved 200 reserved 400 Low Signal IQ 800 Masimo SET This flag means the algorithm is running is full SET mode It requires a SET sensor and needs to acquire some clean data for this flag...

Страница 21: ...THE LIMITS EACH TIME THE PULSE OXIMETER IS USED An audible alarm and a flashing alarm status indicator will occur when an alarm limit is exceeded for greater than five seconds See section 4 Alarm Fea...

Страница 22: ...taneously Return for service 5 a l a r m s a n d m e s s a g e s MESSAGES The Rad 8 will indicate other data or system errors Message conditions for the Rad 8 follow DISPLAY TYPE SOLUTION LEDS FLASH H...

Страница 23: ...s Pulse Rate Accuracy Pulse rate 25 240 bpm No Motion1 Adults Pediatrics Neonate 3 digits Motion2 Adults Pediatrics Neonate 5 digits Low Perfusion3 Adults Pediatrics Neonates 3 digits Resolution Satur...

Страница 24: ...lass B Equipment Classification IEC 60601 1 UL 60601 1 IEC 60601 1 1 Type of Protection Class 1 on AC power Internally powered on battery power Degree of Protection Patient Cable Type BF Applied Part...

Страница 25: ...f monitoring For more information refer to the following table or contact your Sales Representative Use only Masimo SET sensors and sensor cables Select an appropriate sensor apply it as directed and...

Страница 26: ...o L 3 kg 3 3 3 bpm 5 bpm 3 3 bpm 40 kg 2 3 3 bpm 5 bpm 2 3 bpm LNCS NeoPt L 1 kg 3 3 3 bpm 5 bpm 3 3 bpm SENSOR APPLICATION INSTRUCTIONS Unless indicated otherwise in the directions for use reposition...

Страница 27: ...OLUTION DONOTSTER ILIZE ANY MASIMO SENSOR BY IRRADIATION STEAM OR ETHYLENE OXIDE M a s i m o S E T p a t i e n t c a b l e s Reusable patient cables of various lengths are available Only use appropria...

Страница 28: ...ot allow liquids to enter the interior of the instrument Glutaraldehyde Solution Ammonium Chloride Wipe 10 Chlorine bleach in H2O 70 Isopropyl alcohol CAUTIONS DO NOT AUTOCLAVE PRESSURE STERILIZE OR G...

Страница 29: ...ibed tests discontinue its use and correct the problem before returning the unit back to the user Before performing the following tests verify unit is connected to AC power Also disconnect any patient...

Страница 30: ...part of the price paid for this product MASIMO grants to Purchaser a nonexclusive nontransferable license without right to sublicense to use the copy of the incor porated software firmware and documen...

Страница 31: ...vative works based on the written materials without the prior written consent of MASIMO 6 Transfer Restrictions The software firmware is licensed to the Purchaser and may not be transferred to anyone...

Страница 32: ...tion on both the sensors and monitors to ensure accurate pulse oximetry when needed most 2007 Masimo Corporation Masimo Signal IQ Discrete SaturationTransform DST Satshare SET FastSat Rad 8 Radlink Fa...

Отзывы: