Strain Guage Indicator: 409-W

 

REF NO: m59A/om/501 

 

 

Issue NO: 03 

Page 25 of 34 

 

• Sensitivity: 2mV/V 
• Maximum load: 4000.0kg 

 

Set calibration parameters as shown below: 
SPAN = 4000.0 kg 
MV-FS = 20.000 mV 
Zero weight is empty scale weight. 
 
Enter into 

CALZ

 mode with front panel keys, 

 

Zero Calibration: Feed empty scale mV and Press

key 

 

3.  If it does not happen please check mechanical conditions and connection with   

 Indicator. 

 Note: calibrate unit with higher resolution for better result.  

 For Ex.  

With Resolution 4 (9.9999 kg), Resolution 3 (99.999 kg), Resolution 2 (999.99 kg), 
Resolution 1 (9999.9 kg), Resolution 0 (99999 kg). 

 

 

 

 

Содержание 409-W

Страница 1: ...ge 1 of 34 User s Manual 409 W Strain Gauge Indicator Masibus Automation and Instrumentation Pvt Ltd B 30 GIDC Electronics Estate Sector 25 Gandhinagar 382044 Gujarat India Phone 91 79 23287275 79 Fax...

Страница 2: ...Connection diagram for SGI Input 7 6 Configuration Guidelines 8 6 1 Menu Parameter List 8 6 2 Menu Layout for 409 W 12 6 3 Main Menu for 409 W 12 6 4 Calibration menu 14 7 Alarm Operation 15 7 1 Alar...

Страница 3: ...cal Specifications 29 10 1 Input Specification 29 10 2 Output Specifications 29 10 3 General Specifications 30 10 4 Power Supply 31 10 5 Isolation 31 10 6 Communication Specifications 31 10 7 10 7Spec...

Страница 4: ...n is user friendly 1 2 Model and Suffix code Check the model and suffix codes to confirm that the product received is the one which was ordered 1 3 Accessory The product is provided with the following...

Страница 5: ...ment in run mode UP key It is used to increment the parameter for selection Value of Parameter can be incremented by pressing this key If the key is pressed continuously for more than 10 counts change...

Страница 6: ...Strain Guage Indicator 409 W REF NO m59A om 501 Issue NO 03 Page 6 of 34 4 Panel Cutout Dimension Fig 4 1 Panel cut out for 409 W...

Страница 7: ...ator 409 W REF NO m59A om 501 Issue NO 03 Page 7 of 34 5 Sticker details 5 1 Terminal Arrangement Diagram Fig 5 1 Terminal arrangement for 409 W 5 2 Connection diagram for SGI Input Fig 5 2 Connection...

Страница 8: ...T Offset Set offset value for weight 9999 to 9999 0 0 KTARE KTARE Keypad Tare option Set value to enable or disable Key tare 0 Disb 1 En eN ZTARE ZTARE Zero Tare set to clear Tare 0 NO 1 YES NO STEP S...

Страница 9: ...pan for retransmission scaling 200 0 R MAPP R MAPP RET map Set value for retransmission mapping 0 NET 1 Gross 0 Alarm Alarm Logic ATYP ATYP Alarm Type Set which Set Point to shown in SV display in RUN...

Страница 10: ...sec that Relay will wait before getting ON after the ON condition occurs 0 to 9999 sec 0 sec CTRl CTRL Control Relay Select Control Relay Status 0 ON 1 OFF OFF COM Communication SRNO srno Serial No Un...

Страница 11: ...4 0000 00 1 ZERO zero Zero Can be set to any value within the Input Range less the SPAN Value 0 00 SPAN span Span Can be set to any value within the Input Range greater the ZERO Value 200 00 2 mV FS...

Страница 12: ...select 1 15mV for 0 15mV 2 30mV for 0 30mV 3 75mV for 0 75mV Note 3 initial settings Parameters Set value Resolution 0 00 Scaling input zero mV 0 0mV Scaling input Display Zero 0 00 Scaling input span...

Страница 13: ...Strain Guage Indicator 409 W REF NO m59A om 501 Issue NO 03 Page 13 of 34...

Страница 14: ...Strain Guage Indicator 409 W REF NO m59A om 501 Issue NO 03 Page 14 of 34 6 4 Calibration menu Theoretical mV calibration method Sample weight calibration method...

Страница 15: ...d as normal condition in TRIP type 7 3 Latching of ALARM This is used for latching of discrete LEDs and relay status when alarm limit is crossed This option will keep discrete LEDs Relay latched even...

Страница 16: ...ic HL high low AL1 low AL2 high Fig 7 2 AL1 low AL2 High SP2 SP1 If PV SP2 then Relay 2 ON If PV SP2 Hyst2 Relay 2 OFF PV SP1 Relay1 ON If PV SP1 Hyst1 then Relay 1 OFF Depending up on condition set i...

Страница 17: ...8 HH Logic HH high very high AL1 high AL2 very high In this logic if UP Scale condition has been selected than in OPEN sensor condition ALARM 1 and ALARM 2 will be in the ABNORMAL condition and will w...

Страница 18: ...en process value will display OVER UNDER message until value crosses 5 of selected group of mV 15mV or 30mV or 75mV If value of zero and span is less than maximum value of zero and span then process v...

Страница 19: ...e no change in the LED status Alarm AL1 Momentary Alarm when in abnormal condition ACK not pressed Condition Normal Abnormal UP DOWN ACK Normal ACK High Alarm Latch Yes LAMP OFF FLASH FLASH OFF FLASH...

Страница 20: ...LAY OFF ON ON OFF OFF OFF Trip LAMP OFF FLASH OFF OFF FLASH OFF RELAY OFF ON OFF OFF ON OFF High Alarm Latch Yes LAMP OFF FLASH FLASH OFF FLASH OFF RELAY OFF ON ON OFF OFF OFF Alarm Latch No LAMP OFF...

Страница 21: ...OFF RELAY OFF ON OFF OFF ON ON OFF VLOW Alarm Latch Yes LAMP OFF FLASH OFF FLASH STEADY STEADY OFF RELAY OFF ON OFF ON ON OFF OFF Alarm Latch No LAMP OFF FLASH OFF FLASH STEADY OFF OFF RELAY OFF ON O...

Страница 22: ...ndition after abnormal has occurred Latch No RELAY OFF ON ON OFF OFF OFF OFF Trip LAMP OFF FLASH OFF OFF STEADY STEADY OFF RELAY OFF ON OFF OFF ON ON OFF LOW Alarm Latch Yes LAMP OFF FLASH OFF FLASH S...

Страница 23: ...panel keys Display shows Calibration menu 2 Select calibration method Sample weight calibration S WGT a Set resolution enter zero point of weight and span point of weight b Now display indicates Proce...

Страница 24: ...Resolution 40093 Swapped Long 0 4 R W 3 Zero display 40095 Swapped Long 19999 99999 R W 4 Span display 40097 Swapped Long 19999 99999 R W 5 Calibration index 40099 Swapped Long 1 2 R W Table 8 1 Theo...

Страница 25: ...ght is empty scale weight Enter into CALZ mode with front panel keys Zero Calibration Feed empty scale mV and Press key 3 If it does not happen please check mechanical conditions and connection with I...

Страница 26: ...04 Read input registers Use to read programmable registers 05 Force single coil Use to ON OFF single coil 16 Preset Multiple register Use to write programmable register Table 9 1 The error checking fi...

Страница 27: ...Unsigned char 0 4 R W 21 Hysteresis 40009 Unsigned char 0 255 R W 22 Serial number 40010 Unsigned Char 1 247 R W 23 Baud rate 40011 Unsigned char 0 4 R W 24 Alarm logic type 40012 Unsigned char 0 2 R...

Страница 28: ...m status1 Alarm status2 gives status of abnormal condition only Address 1003 1016 for future use only Input type 26 SGI Baud rate 0 4800 1 9600 2 19200 3 38400 Alarm Latch 0 YES 1 NO Alarm sensor 0 UP...

Страница 29: ...gnal Range Accuracy Strain gauge indicator SGI 75mV 19999 to 99999 0 1 of Full span 1 digit Table 14 Digital Input 1 Channel Isolated Non voltage contact input Maximum reverse voltage 6V Maximum Forwa...

Страница 30: ...Plastic Case Color Black Mounting method Panel mounting Dimension 96mm W 48mm H 112mm D Panel Cutout 92mm W 46mm H Weight 260 grams Approx TEMPCO 100 ppm for input to display 150 ppm for retransmissio...

Страница 31: ...1500 V AC for 1 minute Between secondary terminals At least 500 V AC for 1 minute Insulation resistance Between Power supply terminal and ground terminal 500V DC 50 M Primary terminals indicate power...

Страница 32: ...Scalability for SGI input Rx output mapping According to Net and Gross value On demand display value Gross mV Digital Input For Tare Decimal Point User Programmable Step value for display increments...

Страница 33: ...For mA retransmission output short the 2 and 3 pins of jumpers JP8 and JP9 of AO card as shown in figure 7 11 3 Load connection Electrical precautions during use Electrical noise generated by switchi...

Страница 34: ...ss Value 15 00 2 00 Gross Value 30 00 11 5 What is tare weight It is the weight of an empty object With an empty weighing device setting a zero point For the whole system is possible via the input Par...

Отзывы: