Strain Guage Indicator: 409-W
REF NO: m59A/om/501
Issue NO: 03
Page 29 of 34
03
Illegal data value. A value contained in the query data field is not an allowable value for the salve.
06
When Master device write some parameters to Slave device, If slave device busy then it will send 06
code to indicate slave device is busy.
Table 9.3.
10. Technical Specifications
10.1 Input Specification
INPUT TYPES Table:
INPUT
Input Signal
Range
Accuracy
Strain gauge
indicator(SGI)
+ 75mV
-19999 to 99999
0.1 % of Full span ± 1
digit
Table 14.
Digital Input
1-Channel (Isolated) Non- voltage contact
input, Maximum reverse voltage 6V,Maximum
Forward voltage 50V, Capacity 24V DC, 10mA
SPECIFICATIONS:
Sampling Period
4 Sample/Sec
Burn out current
0.5 uA
Noise Rejection Ratio ,Common Mode
Normal Mode
>100 dB (50Hz)
>40 dB (50Hz)
Response time
< 1000mS
Resolution
17-bit
Repeatability
0.05% of FS
Step value, Display Increments
x1, x 2, x 5, x 10
10.2 Output Specifications
RETRANSMISSION OUTPUT
DC Current
0 to 20 mA DC, 4 to 20 mA
DC Voltage
0 to 10 V DC, 0 to 5V DC, 1 to 5V DC.