Strain Guage Indicator: 409-W
REF NO: m59A/om/501
Issue NO: 03
Page 11 of 34
(
brHt
)
10
to
100
F-CNT
(
F-CNT
)
Filter count
IIR Filter is apply for sated count only.
0-500
20
ALPHA
(
ALPHA
)
Alpha
Set alpha for IIR Filter.
0-100
0
M-CNT
(
M-CNT
)
Average Count
Moving Average Filter is apply for sated count only.
0-500
20
PASS
(
pass
)
Password
Set Device Password
0
to
9999
1
CAL**
Calibration
CALIB
(
CALIB
)
Calibration
methods
Set calibration method.
0 :
S-WGT
(Sample weight calibration method)
1 :
TH-mV
(theoretical mV calibration method)
S -WG T
RESO
(
RESO
)
Resolution
Set position of Decimal Point on Display.
0 :
0
1 :
.0
2 :
.00
3 :
.000
4 :
.0000
. 00
1
ZERO
(
zero
)
Zero
Can be set to any value within the Input Range & less the
SPAN Value.
0.00
SPAN
(
span
)
Span
Can be set to any value within the Input Range & greater
the ZERO Value.
200.00
2
mV-FS
(
mv-FS
)
Full span mV
Can be set to any value within the input mV range.
20.000
2
CALZ
(
CALZ
)
Calibrate zero
theory mV
Set above calibration parameters as per requirement
Press DOWN key to calibrate ZERO
Process
value
1
SMP-W
(
SMP-W
)
Sample weight
calibration
Set above calibration parameters as per requirement,
Press DOWN key to calibrate ZERO
press UP key to calibrate SPAN
Process
value
Table 6.1
1
Display only if method is
S-WGT.
2
Display only if method is
th-mV.
Note 1:
For SGI input DI is used to Tare the weight. When 24V Signal is applied
momentarily at the DI terminal, it will tare the weight.