background image

5

3. M 

Ω

INSULATION TEST

1)

Connect the red and black test leads to P + N terminals respectively

2)

Turn dial to M 

Ω

(insulation resistance). Display shows 'Riso' at top left.

3)

Select 'MAN', 'AUTO' or 'TIMER' mode by repeatedly pressing the 
key.  See below for details of individual tests.

4)

Press

repeatedly to select the voltage suitable for the test being

carried out.  50V, 100V, 250V, 500V, 1000V.

5)

Connect the test leads to the conductor on which the insulation test is to
be carried out.

6)

Press yellow          key.  The instrument performs the test as stated below.

At the end of the test, the instrument beeps twice indicating a successful
measurement. The resistance is displayed on the screen. 
If the insulation is very low value the tester emits a long single beep to indicate
that the test voltage fell below the set rated voltage and the measurement was
unsuccessful. 

See technical specification.

MAN:

Minimum test time of 5 or 10 seconds depending on test voltage and
insulation resistance. See table below. If “Go” is pressed and held for
longer than 10 seconds, the instrument continues to measure until the 
“Go” key is released.

AUTO:

The test ends when the measured value becomes stable.

TIMER:

Test duration can be adjusted from 10-999 seconds.  Adjust values

using

or 

This test can be used if a minimum 

measuring time is required.

U

N

/I

Δ

n

FUNC

L

S

U

L

M

GO

100V

250V

500V

1000V

Low resistance

5 seconds

10 seconds

10 seconds

10 seconds

High resistance

5 seconds

5 seconds

5 seconds

5 seconds

Open circuit voltage: Rated voltage test -0% +10%

Short circuit current: <3,0mA @ 500V 

<2,0mA @ 50V, 100V, 250V, 1000V

Test current on load: >1mA @ V

Test

across R 

load

=(1000 x V

Test

)

Ω

If the instrument measures a voltage higher than 30V, the instrument does not
perform the test

RCD

Type of RCD: AC, A

Range of phase to earth voltage: (100 - 265) V

Rated Tripping Current (I

ΔΔ

N): 10mA, 30mA, 100mA, 300mA, 500mA

Test frequency: 50Hz ± 0.5 Hz

TRIPPING TIMES

Test

Range

Resolution

Accuracy

½ I

Δ

N

, I

Δ

N

0-999ms

1ms

±(2%rdg+2dgt)

2 I

Δ

N

0-200ms general

1ms

±(2%rdg+2dgt)

0-250ms selective

1ms

±(2%rdg+2dgt)

5 I

Δ

N

RCD(*)

0-50ms general

1ms

±(2%rdg+2dgt)

0-160ms selective

1ms

±(2%rdg+2dgt)

(*) For RCD type A 500mA, the MAN x5 and AUTO tests are not available. 
The Ramp test measures tripping current

CONTACT VOLTAGE
Measuring range(V)

Resolution(V)

Accuracy

0 - 2Ut Lim

0.1

-0%,+(10%rdg+2dgt)

Ut Lim (UL): 25V or 50V

TRIPPING CURRENT FOR RCD TYPE A AND AC
RCD with I

Δ

N

10mA

Type RCD

Range I

Δ

N(mA)

Resolution(mA)

Accuracy I

Δ

N

AC

(0.5 - 1.4) I

Δ

N

0.1 I

Δ

N

-0%, +10% I

Δ

N

A

(0.5 - 2.4) I

Δ

N

RCD with I

Δ

N

10mA

Type RCD

Range I

Δ

N(mA)

Resolution(mA) Accuracy 

I

Δ

N

AC

(0.5 - 1.4) I

Δ

N

0.1 I

Δ

N

-0%, +10% I

Δ

N

A

(0.5 - 2) I

Δ

N

16

Содержание VERITEST 2240

Страница 1: ...G Motor Maintenance Equipment G Non trip loop testers G Pat testers Accessories G Phase rotation G Proving units G Socket Testers G Thermometers Probes G Test Leads G Voltage Indicators G Specialist...

Страница 2: ......

Страница 3: ...ads 3 2 LOW continuity test 4 3 M insulation test 5 4 RCD test 6 5 Loop Zs IK 9 6 Non Trip Earth Loop Impedance Test Ra15mA 10 1 resolution with PFC calculation 7 Phase Rotation Sequence 11 8 Instrume...

Страница 4: ...rrent Equipment protected by Double Insulation Class II Caution refer to accompanying documents Caution risk of electric shock 10 OPTIONAL ACCESSORIES Infra red data lead VRTL1 Download software VRSOF...

Страница 5: ...219mA the instrument has a stored value for the resistance of the test leads current supplied by instrument during calibration procedure function selected FUNC GO NON TRIP EARTH LOOP TEST RA15MA Meas...

Страница 6: ...ds Use to adjust time up and down respectively At the end of the test if the resistance value is lower than 5 the instrument beeps twice indicating low resistance The resistance value is displayed on...

Страница 7: ...be adjusted from 10 999 seconds Adjust values using or This test can be used if a minimum measuring time is required UN I n FUNC L S UL M GO 100V 250V 500V 1000V Low resistance 5 seconds 10 seconds 10...

Страница 8: ...on display by O B Press the yellow key twice The instrument injects a current in phase with negative half wave of the voltage indicated on display by 180 FUNC UN I n L S S UL M GO GO 9 SPECIFICATION...

Страница 9: ...ct to the voltage Readings below 1 are displayed as 1 PLEASE NOTE MAN x5 and auto test modes are not available for RCD type A 500mA 8 INSTRUMENT CONNECTION TO PC OR A PRINTER The connection from the V...

Страница 10: ...st passed Please refer to the full user manual for tripping times Press to scroll through the test results RAMP The instrument generates a leakage current growing step by step in a given time interval...

Страница 11: ...Connect the mains plug to a 230V 50HZ socket outlet or to the appropriate conductor by means of the croc clips 4 Press key repeatedly to select P PE P PE is shown 5 Press the yellow key once The inst...

Страница 12: ...nt Icc UN Ra When UN phase to earth voltage 127V if 100V Vmeas 150V 230V if 150V Vmeas 265V This test is carried out at 15mA 10 UL M GO 7 PHASE ROTATION SEQUENCE 1 Connect red black and green test lea...

Отзывы: