Series E Ergonomics Force Gauges 

 

User’s Guide 

 
 

 

 

 

21

 

 
10   PASSWORDS 

 
Two separate passwords may be set to control access to the Calibration section and to the menu and 
other keys. To access the passwords setup screen, select 

Passwords

 from the menu. The display 

appears as follows: 
 

 

 

10.1 Calibration Password 

Select 

Calibration

 from the sub-menu. The display appears as follows: 

 

 

 
To set the password, select 

Enabled

 , then 

Set Password

. Use the 

 

and 

 

keys to increment and 

decrement the value, from 0 to 9999. When the desired value has been selected, press 

ENTER

, then 

ESCAPE

 to exit the sub-menu. 

 
10.2 MENU Key Password 

If enabled, every time the 

MENU

 key is selected, a password must be provided. Select 

Menu Key

 from 

the sub-menu. Follow the same procedure as described above.  
 

10.3 Locking Out Other Keys 

Other keys may be locked out individually. Pressing a locked key will prompt the message “KEY 
PROTECTED” and then revert to the previous screen. 
 

10.4 Password Prompts  

If passwords have been enabled, the following will be displayed when pressing the 

MENU

 key or 

accessing the 

Calibration

 section: 

 

 

 
Use the 

 

and 

 

keys to select the correct password, then press 

ENTER

 to continue. 

 
If the incorrect password has been entered, the display appears as follows: 

PASSWORDS 

        

Calibration  

        MENU Key 
             MODE Key 

ZERO Key     
DATA Key 
POWER Key 

 

ENTER PASSWORD 

(0000 – 9999) 

 

5000

 

CALIBRATION 

PASSWORD 

 
  *  

Disabled 

 

     Enabled 
     Set Password 
     (0000 – 9999) 

www.

.com

[email protected]

1.800.561.8187

Содержание ME-100

Страница 1: ...Series E ERGONOMICS FORCE GAUGES User s Guide www com information itm com 1 800 561 8187...

Страница 2: ...des setup safety and operation instructions Dimensions and specifications are also provided For additional information or answers to your questions please do not hesitate to contact us Our technical s...

Страница 3: ...ite software MESURTM gauge software 90 day trial User s Guide 1 E1003 Padded attachment rectangular 1 E1004 Padded attachment curved 1 E1006 Hook 1 E1009 Double handle grip 1 E1000 Carrying case small...

Страница 4: ...rically conductive liquids at all times 3 The gauge should be serviced by a trained technician only AC power must be disconnected and the gauge must be powered off before the housing is opened 4 Alway...

Страница 5: ...ge may occur If the AC adapter is plugged in an icon appears in the lower left corner of the display as follows If the AC adapter is not plugged in battery power drainage is denoted in a five step pro...

Страница 6: ...ent pull back on the lever in the rear of the housing as shown below If the E1011 grip adapter pictured below is used common grips and attachments may be mounted via the integrated 5 16 18F thread 3 1...

Страница 7: ...oses Failure to use a dowel pin properly can result in a hazardous situation 3 2 Installing the USB driver Caution Install the USB driver before physically connecting the gauge to a PC with the USB ca...

Страница 8: ...overload condition is imminent The bar increases either to the right or to the left from the midpoint of the graph Increasing to the right indicates compression load increasing to the left indicates...

Страница 9: ...es between tension and compression directions while configuring set points and other menu items ZERO Zeroes the primary reading and peaks UP Navigates up through the menu and sub menus MENU Enters the...

Страница 10: ...rs as follows One two or none of the set points may be enabled To toggle between the tension and compression directions press the DIRECTION key If two set points have been enabled they are displayed i...

Страница 11: ...the live measured reading 6 2 Peak Compression PC The primary reading corresponds to the peak compression reading observed If the actual force decreases from the peak value the peak will still be ret...

Страница 12: ...0 1 sec increments Averaging Time The time duration of the averaging sequence This setting is ignored when Stop at Force is enabled unless the 300 sec limit is reached Available settings 0 1 300 0 sec...

Страница 13: ...able as shown below 6 5 3 Stop Condition Data capture is automatically terminated when the Stop Condition has been reached Several conditions are available as shown below Function Description Enabled...

Страница 14: ...orage and press ENTER Then exit the menu In the home screen the data record number 0000 appears below the primary reading Press DATA at any time to save the displayed reading The record number will in...

Страница 15: ...ress ENTER to output data and statistics to an external device The display will show SENDING DATA then SENDING STATS then DATA SENT then STATS SENT If there was a problem with communication the displa...

Страница 16: ...baud rate as required for the application It must be set to the same value as the receiving device 8 1 2 Data Format Select the desired data format The display appears as follows Selection Description...

Страница 17: ...e decimal point is dependent of the gauge s resolution Any detected errors are reported back by means of error code 10 illegal command 8 2 DATA Key Functions The DATA key can be configured to perform...

Страница 18: ...ads in each direction 3 To escape the Calibration menu at any time press ESCAPE The display appears as follows Selecting Cancel will revert back to the Calibration setup Selecting Exit w o saving will...

Страница 19: ...ss ENTER 8 The display appears as follows Apply a weight equal to the full scale of the instrument then press ENTER 9 After displaying Please wait the display appears as follows CALIBRATION COMPRESSIO...

Страница 20: ...pression calibration points have been completed the display appears as follows Press ENTER 12 The display appears as follows Reverse the orientation of the load cell shaft by rotating the gauge 180 de...

Страница 21: ...oad is not swinging oscillating or vibrating in any manner Then try again The calibration weight does not match the set value The entered calibration point is too close to the previous point CALIBRATI...

Страница 22: ...ry time the MENU key is selected a password must be provided Select Menu Key from the sub menu Follow the same procedure as described above 10 3 Locking Out Other Keys Other keys may be locked out ind...

Страница 23: ...00 counts or less To access these settings select Automatic Shutoff from the menu The display appears as follows Selection Description Disabled Disable automatic shutoff Enabled Enable automatic shuto...

Страница 24: ...ss ENTER to modify the contrast Select a value from 0 to 25 25 producing the most contrast 11 4 Tones Audible tones can be enabled for all key presses and alerts such as overload set point value reach...

Страница 25: ...ght on up to 7 hours of continuous use Backlight off up to 24 hours of continuous use Measurement units lbF ozF gF kgF N kN depending on model USB output Configurable up to 115 200 baud Safe overload...

Страница 26: ...Settings All disabled Tones Keys Enabled Alerts Enabled Set Points Momentary Automatic Shutoff Enabled Set Minutes 5 Backlight Auto Set Minutes 1 Initial Settings Units lbF Mode Real Time Passwords A...

Отзывы: