Series E Ergonomics Force Gauges
User’s Guide
18
If failed:
6. The following screen appears after the offsets have been calculated:
Attach weight fixtures (brackets, hooks, etc), as required. Do not yet attach any weights or apply
any calibration loads. Press
ENTER
.
7. The display appears as follows:
Optionally exercise the load cell shaft several times (at full scale, if possible), then press
ENTER
.
8. The display appears as follows:
Apply a weight equal to the full scale of the instrument, then press
ENTER
.
9. After displaying “Please wait…” the display appears as follows:
CALIBRATION
COMPRESSION
Gain adjust
Apply full scale load
100.00 lbF +/-20%,
then press ENTER.
CALIBRATION
COMPRESSION
Attach necessary
weight fixtures,
then press ENTER.
CALIBRATION
COMPRESSION
Optionally exercise
sensor, then press
ENTER.
CALIBRATION
OFFSET
Sensor failed
Analog failed
CALIBRATION
OFFSET
Sensor passed
Analog passed
CALIBRATION
OFFSET
Please wait…