9 Messages
|
ROTAFLOW with ICU Kit
|
STOP ???
Yes
Yes
Stopped by an intervention not yet identified.
Possible causes:
Communication error if monitoring system has
been switched off
Immediately after reprogramming of ROTAFLOW
Console
STOP BUB
Yes
Yes
Stopped by bubble monitoring
STOP LEV
Yes
Yes
Stopped by level monitoring
STOP PRX
Yes
Yes
Stopped by pressure monitoring
[X]
: Pressure monitoring 1 to 4
STOP-INP
Yes
Yes
Monitoring intervention
TEMP B
⇑⇑
No
No
Temperature of battery > 50°C
TEMP C
⇑⇑
No
No
Temperature of electronics > 60°C
TEMP M
⇑⇑
No
No
Temperature of ROTAFLOW Drive > 50°C
UNEXPECT
Yes
Yes
Software
Notes: Temperatures
If the temperature of the electronics exceeds 45°C, the ventilator switches to
full speed.
If the ROTAFLOW System stops because the temperature is too high, it
cannot be started again until the temperature has fallen.
No temperature messages should appear under normal operating conditions.
However, if this does happen, contact MAQUET Service.
Message
Acoustic
alarm
Pump
stop
Cause/Error source