4 Operation
|
ROTAFLOW with ICU Kit
|
4.7
MODES “ART” AND “ART PULS”: INTERVENTIONS
In the modes “ART” and “ART PULS” the following interventions can be used.
Selection is made on the heart-lung machine.
Stop when upper pressure limit is exceeded
Stop when bubbles are detected
Stop when the level of liquid in the reservoir is low
Monitoring is performed via the monitoring systems of the heart-lung machine
(
Ö
User’s Manual for heart-lung machine).
The status display indicates whether the ROTAFLOW System was stopped by an
intervention. The heart-lung machine emits an acoustic signal.
STOP BUB
Stopped by bubble monitoring
STOP LEV
Stopped by level monitoring
STOP PRx
Stopped by pressure monitoring x (x = 1 to 4).
WARNING! Pump stop with communication error with heart-lung machine
If the monitoring system of the heart-lung machine is switched off during
operation or a communication error occurs, the ROTAFLOW System will stop,
and the relevant intervention LED will flash. An acoustic alarm is emitted and the
status display flashes
[NO COMM.].