3 Connection and Installation
|
ROTAFLOW with ICU Kit
|
1
Connect the inlet tube to the inlet in
the centre of the ROTAFLOW
Centrifugal Pump.
2
Connect the outlet tube to the outlet
on the side of the ROTAFLOW
Centrifugal Pump.
3
Apply ultrasonic contact cream to
both sides (left and right) around the
outlet of the ROTAFLOW Centrifugal
Pump. Both sides must be
completely covered with the
ultrasonic contact cream.
4
Insert the ROTAFLOW Centrifugal
Pump into the ROTAFLOW Drive.
Check that the ROTAFLOW
Centrifugal Pump is positioned under
the holder of the ROTAFLOW Drive.
5
Close the lock.