3 Connection and Installation
|
ROTAFLOW with ICU Kit
|
3.5
OPERATION AS A STAND-ALONE DEVICE: INSTALL LEVEL
SENSOR
Note
In the “ART” and “ART PULS” modes, the level is monitored by the heart-lung
machine (
4.7 “Modes “ART” and “ART PULS”: Interventions”, page 38 and the
Ö
User’s Manual for heart-lung machine).
WARNING!
Attach the level sensor pad where the wall of the reservoir is between 2 and
5 mm thick. Do not attach the level sensor pad to a greasy, dusty or uneven
surface or to a label coated with metal. Do not attach anywhere where blood
flows along the wall as a thin film of blood.
Try to prevent liquids from coming into contact with the level sensor pad. If a
level sensor pad becomes wet, replace it immediately with a dry one.
Do not use level sensor pads which are damaged, have become detached or
do not have an even contact surface.
Do not touch the contact surface.
Do not re-use level sensor pads as this might impair sensitivity.
There must be at least 30 mm of liquid between the level sensor and the
opposite wall.
With reservoirs containing metal heat exchangers, the distance between the
reservoir wall and the metal spiral must be at least 40 mm.
Do not attach more than one level sensor to the reservoir.
1
Remove the protective paper and
attach the level sensor pad onto the
reservoir horizontally. The arrow
must be pointing to the right at the
desired height.
2
Attach the level sensor to the level
sensor pad.