ROTAFLOW with ICU Kit
|
3 Connection and Installation
|
29
|
ROTAFLOW with ICU Kit | User’s Manual | English | 1.0
Copyright MAQUET Cardiopulmonary AG, Hirrlingen
3.7
SWITCH ON ROTAFLOW CONSOLE, SELF-TEST
The ROTAFLOW Console will automatically perform a self-test after being
switched on.
4
At the beginning of the self-test check that all displays and LEDs are lit up and
that an acoustic alarm is emitted.
The ROTAFLOW Console shows various messages during the self-test (
“Messages during self-test”, page 49).
5
Check whether the displays show the following messages at the end of the self-
test:
[–OK– EX.X PUMP OK]
On the display, [X.X] will indicate the software version.
6
If the displays then show the message
[VALVE?]
, you need to confirm that a
valve or other suitable systems or methods are available (e.g., two tube
clamps) to prevent a reverse flow. Press the “Clamp” button.
WARNING!
When using as a stand-alone device, switch on the mains power switch at rear of
ROTAFLOW Console.
Otherwise the ROTAFLOW Console will default to battery operation after being
switched on, thus discharging the batteries.
1
Open the cover of the on/off switch
on the front.
2
Press the on/off switch.
3
Close the cover again.