Mantracourt Electronics Limited
– FSU User Manual
13
System Zero
System Zero is used to remove permanent values from an installed system such as the weight of a vessel or weighing
platform to give a zero Gross value.
The value shown on this page is subtracted from the measured value.
You can either type a value in manually or click the
Zero Now
button to automatically fill in the value to give a zero
reading based on the current input.
This feature can be hidden on the
Lock Page
.
Auto Calibration
The automatic calibration requires that you apply two known (ascending) inputs in sequence. Just enter the required
engineering unit value for the two inputs in the fields marked 1 and 2.
Next, apply the low input and allow a settle time. Click the ‘
Acquire’
button then apply the higher input. After
another settle time click the second ‘
Acquire’
button and the calibration is complete.